Health O Meter 500KL operation manual Assembly Instructions, Power Source, Calibration

Page 4

OPERATING INSTRUCTIONS

MODEL 500KL

OVERLOAD

OVEr

UndEr

6.2Overload indication. When the scale is overloaded, “OVER” will shows.

6.3Negative weight. When there is a negative weight value, “UNDER” shows and the scale will correct this error by rezero-setting.

7. POWER SOURCE

ZERO

0.0

KG

7.16*AA batteries or a 12V adapter. When the adapter is used, the battery icon will not be displayed

C

8.CALIBRATION

1.Press UP and Down at the same time, then press ON/OFF to start the scale. LCD will displays ”CAL”, then returns to “0.0 LB”

2.Press DOWN key, it displays”-100”, then put on 100lb weight. After display data becomes stable, Press DOWN key, it displays “CAL” then ”-200”,

3.Put on 200lb weight. After display data becomes stable, press DOWN key, it displays “CAL” then ”-300”.

4.Put on 300lb weight. After display data becomes stable, press DOWN key,it displays "CAL" then "-400".

5.Put on 400lb weight. After display data becomes stable, press DOWN key, it displays "CAL" then "-500"

6.Put on 500lb weight. After display data becomes stable, press DOWN key, it displays "CAL" then auto off automatically.

ASSEMBLY INSTRUCTIONS

1.Connect Lower Pillar (B) to Base (C) using 4 Screws (D).

2.Connect Upper Pillar (A) to Lower Pillar (B).

3.Connect Upper Height Rod (E) to Lower Height Rod (F).

4.Place assembled height rod into slot (H) as shown. Pinch stops (G) to allow entry.

H

A

E

F

B

G

C

D

4

Image 4
Contents Digital Physician Scale Table of Contents Definition of SymbolsTo Start the Scale for Weight Measurement Operating InstructionsButtons BMI CalculationPower Source Assembly InstructionsCalibration Limited Warranty What does the Warranty Cover?Mode D’EMPLOI Definition DE Symbole Precaution ET PreventionTable DES Matieres Nstructions D’U Tilisation TouchesBMI Calcul Valoir FonctionInstructions D’UTILISATION AssemblagePrécautions ET Avertissements Source D’ALIMENTATION DES BatteriesGarantie Limitée D’UN AN Interdit à l’Usage CommercialManual DE Operación Índice Precaución Y AdvertenciaDefinicion DE Simbolos Modelo 500KLInstrucciones DE Operación BotonesPara Activar LA Balanza Y Determinar EL Peso Cálculo DEL Índice DE Masa CorporalFuente DE Alimentación ArmadoCalibración Garantía Limitada ¿Qué Cubre la Garantía?