CALIBRATION
When to Calibrate
Calibration may be required when the scale is initially installed, if it is moved a substantial distance, or in accordance with local regulations. This is necessary because the weight of a mass in one location is not necessarily the same in another location. Also, with time and use, mechanical deviations can occur.
The Position of the [UNITS] key for the Calibration
The (k)g model has no
[UNITS] key, but the key for the calibration is located under the overlay.
Calibration Using a Calibration Weight
Calibrate the scale with the following calibration weight.
1000g ± 0.1g | 2000g ± 0.2g | ||
5000g ± 0.5g | 5000g ± 0.5g | ||
10kg ± 1g | 20kg ± 2g | ||
|
| 30kg ± 2g |
1.Enter calibration mode.
Press the [ON/OFF] key to turn the power off.
While pressing both the
2.Zero calibration
Press the
Cal0
will be displayed.
When you have proceeded to the display other than Cal0 the [ON/OFF] key to turn the power off and start over again.
, press
Wait for the stable indicator to be displayed and press the
3.Span calibration
Place the weight at the center of the platform for accurate calibration.
Wait for the stable indicator to be displayed and press the
-5 -