Lumex Syatems LF1050 Checking the Battery, Charging the Batteries, Important Notes on Charging

Page 12

CHECKING THE BATTERY

1.Check batteries by pressing the Battery Testing Button (blue circle with a battery sign) on the control box.

2.Batteries are fully charged when all lights on the Battery Indicator are "ON".

3.Charge batteries when Indicator shows only one “GREEN” light.

4.Do not use the lift when no “GREEN” light is shown. Charge the batteries immediately.

5.Replace batteries when frequent charging is needed. (Reorder #: DPL650BAT)

CHARGING THE BATTERIES

1.Ensure the power is switched "ON" (the red “RESET” button should be up).

2.Insert charging plug into charging port on the control box.

3.Plug charger to power supply.

4.All lights of battery indicator should be "ON" while charging.

5.It takes approximately 2-3 hours to fully charge the batteries from one green light.

6.It takes approximately 7-9 hours to fully charge the batteries from the red light.

7.Do not leave the batteries on charging for an extended period of time. This will shorten the life of the batteries.

8.Do not let the batteries run down to the last red light. This might shorten the life of the batteries or damage the batteries.

9.Unplug the charger first before using the lift.

THE LF1050 PATIENT LIFT WILL NOT OPERATE WHILE CHARGING.

IMPORTANT NOTES ON CHARGING

1.Push lift to charging location and charge the batteries with the charger provided. Avoid unplugging hand control and motor from control box. Frequent plug/unplug of the hand control and motor into/from the control box may damage the control box.

2.Do not charge batteries over 12 hours.

Know What The Battery Indicator Lights Are Telling You

Red

Amber

Green

Green

Green

 

 

 

 

 

 

 

ON

ON

ON

ON

ON

FULLY CHARGED

 

 

 

 

 

 

ON

ON

ON

OFF

OFF

NEEDS CHARGE

 

 

 

 

 

 

ON

ON/OFF

OFF

OFF

OFF

CHARGE IMMEDIATELY. DON’T USE THE LIFT

 

 

 

 

 

 

OFF

OFF

OFF

OFF

OFF

BAD BATTERIES OR BAD CONNECTION

 

 

 

 

 

 

Warning Buzzers

The control box has a buzzer that provides you with information on the status of the batteries and the lift. To stop the buzzer, push down the RESET button or press the Battery Indicator.

BUZZER TYPE

INDICATION

NOTE

2 beeps/sec

Low Battery (red / amber lights on)

Charge batteries immediately

 

 

 

3 beeps/sec

Low Battery (red light on)

Charge batteries immediately

 

 

 

6 beeps/sec

Very Low Battery (no light on)

Charge batteries immediately

 

 

 

3 beeps every

Bad Battery (no light on)

Replace batteries

other second

 

 

 

 

 

3 long beeps

Overloaded. Batteries are charged

Exceeded maximum load capacity

but lift does not lift.

 

 

12

LF1050 Patient Lift Owner’s Manual

Image 12
Contents Patient Lift Page Contents Introduction Special NotesMaintenance Safety PrecautionsPage Features Handling ProceduresUnpacking InspectionAssembly Instructions Page Operating Instructions Preparation Before LiftingLift and Transfer From Bed Transfer From Wheelchair Checking the Battery Know What The Battery Indicator Lights Are Telling YouCharging the Batteries Important Notes on ChargingCare and Maintenance CleaningLubrication Checking for damage and wearMaintenance Schedule Troubleshooting Symptom Possible Cause ActionSpecifications Limited Warranty Warranty Shall be Limited to SuchForegoing Warranty is Exclusive Page A., Corporate Headquarters