TROUBLESHOOTING
Warning: If you experience a problem with your lift and are unable to service it yourself, contact your Lumex dealer.
Symptom | Possible Cause | Action | |
|
|
| |
| Battery discharged | Charge battery | |
Control box is not responding and | Battery charger defective | Replace battery charger | |
relay is not clicking | Battery defective | Replace battery | |
| Hand control defective | Replace hand control | |
|
|
| |
| Actuator plug not properly | Push actuator plug firmly into control | |
Actuator does not run, | seated | box | |
but relay is clicking | Actuator defective | Replace actuator | |
| Control box defective | Replace control box | |
|
|
| |
Actuator goes up or down, but not | Hand control defective | Replace hand control | |
both | Control box defective | Replace control box | |
|
|
| |
Cannot lift patient up and battery | Battery discharged | Charge the batteries | |
indicator shows no GREEN light | |||
|
| ||
|
|
| |
|
| Check the connection between the | |
Lift does not work and the battery |
| actuator and the control box | |
indicator shows full charge |
| Check connection between Pendant & | |
|
| control box | |
|
|
| |
|
| Make sure the emergency stop button | |
Lift does not work and the battery |
| is released | |
indicator shows no light |
| Check the connection of the batteries | |
|
| in control box | |
| Check socket for pendant | ||
lowering patient |
| Check pendant cable | |
|
|
| |
Battery Indicator shows no light |
| Check the emergency stop button is | |
| released. | ||
when charger is connected to |
| ||
| Check charger plug for damage. | ||
power supply |
| ||
| Check charger cable for damage. | ||
|
| ||
|
|
| |
Lift does not work. Battery indicator |
|
| |
shows full charge. Notice “click” |
|
| |
sound from control box when UP |
| Actuator problem. | |
or DOWN button on hand control |
|
| |
was pressed. |
|
| |
|
|
|
LF1050 Patient Lift Owner’s Manual | 15 |