Fluke 1523, 1524 manual Safety Information

Page 10

1523, 1524 Thermometer Readout

Safety Information

1.3Safety Information

The Reference Thermometer is designed in accordance with EN 61010-1 {2nd Edition}, and CAN/CSA 22.2 No 61010.1-04. Use this instrument only as specified in this manual, otherwise the protection provided by the instrument may be impaired.

AWarning identifies conditions and actions that pose hazard(s) to the user; a Caution identifies conditions and actions that may damage the instrument being used.

International symbols used on the reference thermometer and in this manual are explained in Table 1 on page 4.

1.3.1 Warning

To avoid possible electric shock or personal injury:

●●Do not use the reference thermometer in environments other than those listed in the user’s guide.

●●Do not use the reference thermometer for any application other than that which is specified. The instrument was designed for temperature measurement and calibration. Any other use of the instrument may cause unknown hazards to the user.

●●If the reference thermometer is used in a manner not in accordance with the equipment design, the operation and the protection provided by the instrument may be impaired. In addition, safety hazards may arise.

●●Do not apply more than the rated voltage, as marked on the reference thermometer, between the inputs, or between any input and earth ground (30 V, 24 mA max all terminals).

●●Follow all equipment safety procedures.

●●Calibration equipment should only be used by trained personnel.

●●The reference thermometer is intended for indoor use only.

●●Before you use the instrument, inspect the case. Look for cracks or missing plastic. Pay particular attention to the insulation surrounding the connectors. Do not use the reference thermometer if it

2

Image 10
Contents 1523 Fluke Corporation, Hart Scientific Division Limited Warranty & Limitation of Liability Fluke Corporation, Hart Scientific Division Table of Contents Before You Start Quick StartFigures Tables Page Introduction Standard EquipmentSafety Information Before You Start Symbol Description Using Clamp-On Ferrites CE CommentsEmissions Testing Low Voltage Directive SafetyAuthorized Service Centers Fluke Intl Corporation ChinaPage Setup Quick Start12V DC Stats Hold Stats 1st Press Enters Recall menu, 2nd Press Exits Recall Menu 1523 Menu 1523 Menu 1523, 1524 Thermometer Readout Setup 1524 Key Functions + Enter 1524 Menu 1524 Menu Quick Start 1523, 1524 Thermometer Readout Quick Start Specifications MV to 75 mV 001 mV Excitation Current 36 C Accuracy ± C 4 Wire Probe