Fluke 1524, 1523 manual MV to 75 mV 001 mV

Page 33

Quick Start

Specifications

Power requirements

 

3 AA alkaline batteries

 

 

 

12 V dc universal power supply

 

 

 

 

 

 

Size

 

96 x 200 x 47 mm

 

 

 

(3.75 x 7.9 x 1.86 inches)

 

 

 

 

 

 

Weight

 

0.65 kg (1.4 lb)

 

 

 

 

 

 

 

Safety

 

EN 61010-1:2001, CAN/CSA C22.2 No.

 

 

 

61010.1-04

 

 

Environmental conditions for all specifications: 13 °C to 33 °C

 

 

 

 

 

Table 7 Millivolt Measurement

 

 

 

 

 

 

Range

Resolution

Accuracy

 

 

 

 

 

–10 mV to 75 mV

0.001 mV

± (0.005 % + 5 V)

 

 

 

 

 

 

Temperature Coefficient ( –10 °C to 13 °C, +33 °C to 60 °C):

± (0.001 %/°C + 1 V/°C)

Table 8 Ohms Measurement, RTDs

 

Accuracy ±

Ohms Range

4 Wire

 

 

0 Ω to 400 Ω

± (0.004 % + 0.002 )

 

 

Temperature Coefficient ( –10 °C to 13 °C, +33 °C to 60 °C):

0.0008 %/°C + 0.0004

Excitation Current: 1 mA

25

Image 33
Contents 1523 Fluke Corporation, Hart Scientific Division Limited Warranty & Limitation of Liability Fluke Corporation, Hart Scientific Division Quick Start Table of Contents Before You StartFigures Tables Page Standard Equipment IntroductionSafety Information Before You Start Symbol Description CE Comments Using Clamp-On FerritesLow Voltage Directive Safety Emissions TestingFluke Intl Corporation China Authorized Service CentersPage Quick Start Setup12V DC Stats Hold Stats 1st Press Enters Recall menu, 2nd Press Exits Recall Menu 1523 Menu 1523 Menu 1523, 1524 Thermometer Readout Setup 1524 Key Functions + Enter 1524 Menu 1524 Menu Quick Start 1523, 1524 Thermometer Readout Quick Start Specifications MV to 75 mV 001 mV Excitation Current 36 C Accuracy ± C 4 Wire Probe