1523, 1524 Thermometer Readout
Specifications
Table 9 Ohms Measurement, Thermistor
Ohms Range | Accuracy ± Ω, 4 Wire |
|
|
200 Ω to 50 kΩ | ± (0.01 % + 0.5 Ω) |
|
|
50 kΩ to 500 kΩ | ± (0.03 %) |
|
|
Temperature Coefficient (
0.002 %/°C + 0.1 Ω (0 Ω to 50 kΩ)
0.06 %/°C + 0.1 Ω (50 kΩ to 500 kΩ)
Excitation Current:
10 ∝A (0 Ω to 50 kΩ)
2 ∝A (50 kΩ to 500 kΩ)
Equivalent temperature accuracies derived from primary specifications (Ω, mV)
Table 10 Temperature, Thermocouples
|
|
| Measure |
Type |
| Range | Accuracies |
|
|
|
|
B | 600 | °C to 800 °C | 0.85 °C |
| 800 °C to 1000 °C | 0.68 °C | |
| 1000 | °C to 1800 °C | 0.57 °C |
|
|
|
|
C | 100 | °C to 550 °C | 0.32 °C |
| 550 °C to 2300 °C | 0.71 °C | |
|
|
| |
E | 0.52 °C | ||
| 0 °C to 950 °C | 0.22 °C | |
|
|
| |
J | 0.52 °C | ||
| 0 °C to 1200 °C | 0.23 °C | |
|
|
| |
K | 0.61 °C | ||
| 0 °C to 1370 °C | 0.24 °C | |
|
|
|
|
26