Fluke 1523, 1524 manual Accuracy ± C 4 Wire Probe

Page 36

1523, 1524 Thermometer Readout

Specifications

Table 11 Temperature, RTD Ranges,

and Accuracies (ITS-90)

Accuracy ± °C 4 Wire Probe

± 0.011 at –100 °C

± 0.015 at 0 °C

± 0.019 at 100 °C

± 0.023 at 200 °C

± 0.031 at 400 °C

± 0.039 at 600 °C

Resolution: 0.001°C (0.001°F)

Table 12 Temperature, Thermistor

Accuracy ± °C

± 0.002 at 0 °C

± 0.003 at 25 °C

± 0.006 at 50 °C

± 0.014 at 75 °C

± 0.030 at 100 °C

Resolution: 0.001 °C (0.001 °F)

Based on a 10 kΩ (at 25 °C) thermister with a beta value of 4000 Ω. See technical manual for details.

28

Image 36
Contents 1523 Fluke Corporation, Hart Scientific Division Limited Warranty & Limitation of Liability Fluke Corporation, Hart Scientific Division Table of Contents Before You Start Quick StartFigures Tables Page Introduction Standard EquipmentSafety Information Before You Start Symbol Description Using Clamp-On Ferrites CE CommentsEmissions Testing Low Voltage Directive SafetyAuthorized Service Centers Fluke Intl Corporation ChinaPage Setup Quick Start12V DC Stats Hold Stats 1st Press Enters Recall menu, 2nd Press Exits Recall Menu 1523 Menu 1523 Menu 1523, 1524 Thermometer Readout Setup 1524 Key Functions + Enter 1524 Menu 1524 Menu Quick Start 1523, 1524 Thermometer Readout Quick Start Specifications MV to 75 mV 001 mV Excitation Current 36 C Accuracy ± C 4 Wire Probe