Abbott Diabetes Care Blood Glucose Monitor Using Your Meter’s Lights, Using the Display Backlight

Page 38

Appendices

1. Using Your Meter’s Lights

Depending on what mode your meter is in, pressing the light button ( ) lights either the display backlight or the test light.

Using the Display Backlight

Pressing the light button powers on the display backlight in all modes except when applying a sample (for example, clock, memory and set).

The display backlight goes out when you press the light button ( ) again or when the meter is inactive for 10 seconds.

Using the Test Light

Pressing the light button while the Blood Drop and Test Strip symbols appear on the display screen powers on the test light. The test light goes out:

When you press the light button again.

When the meter receives the correct amount of blood for a test (the test strip is filled).

After 60 seconds of no activity.

When a test strip is removed from the meter.

35 Call Customer Care Toll Free at 1-888-522-5226

www.AbbottDiabetesCare.com

Image 38
Contents Owner’s Booklet Table of Contents Appendices Test On or Off Your Fingers How the FreeStyle Flash System Should Be Used Important Health-Related Information Data on file testing was done on forearm FreeStyle Flash Meter and FreeStyle Test Strip Getting Started with the FreeStyle Flash MeterFreeStyle Flash System FreeStyle Flash Meter and How It Works Configure Button Test Strip PortFreeStyle Flash Meter Display Check Owner’s Booklet SymbolControl Solution Test Result Symbol Message AreaSystem Check Screen Battery SymbolMemory Mode Symbol Test Result AreaCheck and Set Test Strip Code Symbol Blood Drop and Test Strip SymbolsDate Setup Symbol Thermometer SymbolSymbol Day Average SymbolClock Symbol Sound Level SymbolsFreeStyle Test Strip Test Strip Does the WorkImportant Test Strip Information FreeStyle Control Solution All About FreeStyle Control SolutionWhen should I perform a control solution test? Ensuring Accurate Control Solution Results Where can I obtain FreeStyle Control Solution? Performing a Control Solution Test Top BottomUsed only once FreeStyle Lancing Device Twist Off the Rounded Top Using the Lancing DeviceInsert Lancet Cock the Handle Replace CapSet the Lancing Level Lancing Lancing Your Forearm, Upper Arm, Hand, Thigh, or CalfLancing Lancing Your FingerPotential Bruising MaintenanceRemoving the Lancet Testing Your Blood Glucose Blood Glucose Testing Components Setting the Test Strip Code Performing a Blood Glucose TestCode Number Filling the Test Strip with Blood Waiting for Result Do not lift the test strip up until you hearLO Readings LO Reading with SymptomsLO Reading without Symptoms LO ReadingHI Readings HI Reading with SymptomsHI Reading without Symptoms HI ReadingUsing the Test Light Using Your Meter’s LightsUsing the Display Backlight Using Alarms Setting Up Your Meter Meter Screen DisplaysSetting the Alarm Setting the Sound Setting the Time Setting the Date Setting the Unit of MeasurementSoftware Version Display Day Average Viewing Your Meter MemoryGetting In and Out of Memory Blood Glucose Test Control Solution TestWorking the Memory Special Notes Transferring DataError Messages ErrorError Error Problem with test strip Meter malfunction Batteries Replacing the Batteries Page Troubleshooting Call Customer Care Cleaning the Meter Caring for Your MeterMeter Specifications Index Page Call Customer Care Toll Free at Page Call Customer Care Toll Free at Page Call Customer Care Toll Free at Page Meter Serial No Covered by one or more of the following U.S. Patents

Blood Glucose Monitor specifications

Abbott Diabetes Care has established itself as a leader in the field of blood glucose monitoring, providing innovative solutions for individuals managing diabetes. Among its range of products, the FreeStyle line of blood glucose monitors stands out due to its advanced technology and user-friendly design.

One of the key features of the Abbott FreeStyle Blood Glucose Monitor is its accuracy. The device utilizes advanced algorithms and state-of-the-art biosensors to deliver precise readings, ensuring that users can make informed decisions regarding their diabetes management. With a small sample size requirement, users can obtain their glucose levels with minimal discomfort, making the monitoring process less intrusive.

Another significant aspect of the Abbott monitors is their fast response time. Users can receive their blood glucose results in just a few seconds, allowing for immediate feedback on their glucose levels. This rapid response is crucial for making quick adjustments to insulin doses or dietary choices, empowering individuals to take charge of their health in real-time.

The integration of smart technology is another highlight of Abbott's blood glucose monitors. Many models are equipped with Bluetooth connectivity, enabling seamless communication with compatible smartphones and health apps. This feature allows users to track their glucose levels, analyze trends, and share results with healthcare teams, creating a comprehensive approach to diabetes management. The FreeStyle app offers reminders, data logging, and insights, enhancing the overall user experience.

The monitors are also designed with user convenience in mind. Features such as large, easy-to-read displays and intuitive interfaces make them accessible to a broad range of users, including those who may be elderly or have visual impairments. Additionally, the compact design of the monitors makes them portable, enabling users to carry them easily.

Safety and sanitation are paramount in diabetes care. Abbott has incorporated features such as auto-coding and smart test strip technology to minimize the risk of error. The test strips are designed to be inserted easily and require no manual batch coding, reducing the potential for mishaps in testing.

Overall, the Abbott Diabetes Care Blood Glucose Monitors represent a blend of precision, technology, and user-centric design. With their focus on accuracy, speed, and connectivity, these monitors play a critical role in helping individuals effectively manage their diabetes, leading to improved health outcomes and enhanced quality of life.