GETTING YOUR InDuo™ SYSTEM READY FOR USE 15
æCaution:
•Use only OneTouch® Ultra Control Solution.
•Check the expiration date on the control solution vial. Do not use if expired.
•Store control solution tightly closed at temperatures below 30°C (86ºF).
•Do not refrigerate control solution.
•The control solution, the InDuo™ meter, and the OneTouch® Ultra Test Strips should be at room temperature
•Shake the vial, discard the first drop of control solution and wipe off the dispenser tip with a clean tissue to obtain a good sample and an accurate result.
•Use only for 3 months after the first opening. Record the discard date (date opened plus 3 months) on the control solution vial. Discard after 3 months.
•The control solution range printed on the OneTouch® Ultra Test Strip vial is for OneTouch® Ultra Control Solution only. It is used to test meter and OneTouch® Ultra Test Strip performance. It is not a recommended range for your blood glucose level.