Health O Meter 2450KL operation manual Troubleshooting, Symptom Possible Cause Corrective Action

Page 19

TROUBLESHOOTING

Refer to the following instructions to check and correct any failure before contacting service personnel.

SYMPTOM

 

POSSIBLE CAUSE

 

 

CORRECTIVE ACTION

 

 

 

 

 

 

Scale does not turn on

1.

Dead Battery

1.

 

Replace batteries

 

2.

Faulty electrical outlet

2. Use a different outlet

 

3.

Bad power supply

3.

 

Replace adapter

Questionable weight or

1.

External object

1.

 

Remove interfering object

the scale does not zero

 

interfering with the scale

 

 

from the scale

 

2. The display did not show

2.

Ask the patient to get off

 

 

”0.0” before weighing

 

 

the scale, zero the scale

 

 

 

 

 

and begin weighing

 

 

 

 

 

process again

 

3. Scale is not placed on a

3.

Place the scale on a level

 

 

level floor

 

 

floor and begin weighing

 

 

 

 

 

process again

 

4.

Scale is out of calibration

4.

Check weight with known

 

 

 

 

 

weight value

 

5.

Improper Tare

5.

Place the item to be tared

 

 

 

 

 

on the scale ramps. Press

 

 

 

 

 

REWEIGH. Once the

 

 

 

 

 

weight of the item is

 

 

 

 

 

displayed, press TARE.

 

 

 

 

 

Place the patient and the

 

 

 

 

 

tared item back on the

 

 

 

 

 

scale. Press REWEIGH

 

 

 

 

 

again.

Weighing is performed but

The patient is not sitting still

Ask the patient to be still or

the display shows “weigh”

 

 

you can change to live weight

and “reweigh” every few

 

 

setting

seconds; the weighing

 

 

 

 

 

process takes too long and

 

 

 

 

 

no weight is displayed.

 

 

 

 

 

The display shows

The load on the scale

Remove the excess weight

“Overload” message

exceeds the capacity

and use the scale according to

 

(600 Lbs)

its limits

The display shows

The batteries are empty

Replace batteries according to

“LOW BAT” message

 

 

instructions

The display shows

There is a problem with one

Check load cell cable

“Load Cell Error” message

or more load cells or the load

connection at the display and

 

cell cable is disconnected.

platform assembly ports. If the

 

 

 

problem is not corrected, refer

 

 

 

to qualified service personnel

 

 

 

to replace the defective load

 

 

 

cell

 

18

Image 19 Contents
PRO Plustm Portable Wheelchair Scale Table of Contents Scale Specifications SpecificationsGeneral Scale Opening and Assembly Instructions SET UPSET UP Replacing Batteries Preparing the Scale for UseOperation Instructions Description Function Menu Weight Display ModeOnly applies to pound values, not the metric values Removing the Tare Tare WeightData Management Automatic TareEnter Scale Settings Operation Instructions System Settings System Test User Settings Operation Instructions Disconnect the scale from the AC power source MaintenanceMaintenance CleaningSymptom Possible Cause Corrective Action TroubleshootingOperator Action Display Calibration PathExploded View of Scale continue Description Qty Parts ListTo get Warranty Service WarrantyLimited Warranty What does the Warranty Cover?