Health O Meter 2450KL operation manual Calibration Path, Operator Action Display

Page 20

CALIBRATION PATH

The calibration is performed in Kg or Lb, according to the units of measure used upon entering into the calibration path.

 

Operator Action

 

 

 

Display

1.

Press ON/OFF button and immediately press and

 

Calibration

 

 

 

 

 

hold HOLD/RELEASE button for 3-4 seconds

 

Process

 

 

 

 

2.

Press ENTER

 

Enter load weight

 

 

 

 

 

600.0

 

 

 

3.

Using cd keys and/or the keypad, set the

 

Zero calibration

 

 

 

calibration load to at least 200 lbs (for best results

 

Please clear the

scale

 

and greater accuracy, use a 600 lb weight). Press

 

 

 

 

 

 

 

 

ENTER

 

 

 

 

 

 

 

4.

Please clear the scale platform and press ENTER

 

Zero calibration

 

 

 

 

 

Please wait

 

 

5.

Please wait without touching or shaking the scale

 

Weight calibration

 

 

until the zero calibration process is finished

 

Put: XXX.X

 

6.

Load the scale platform with the required weight

 

Weight calibration

 

 

and press ENTER

 

Please wait

 

7.

Please wait without touching or shaking the scale

 

Calibration

 

 

until the calibration process is finished

 

factor: X.XXXXX

8.

Press ENTER

 

Calibration

 

 

 

 

Please clear the scale

9.

Please remove weight from the scale platform

 

Rebooting

 

 

 

 

Please wait

10.

Please wait until the scale resumes normal

 

Health O Meter

 

operation

 

Pro Plus

EXPLODED VIEW OF SCALE

19

Image 20
Contents PRO Plustm Portable Wheelchair Scale Table of Contents General SpecificationsScale Specifications SET UP Scale Opening and Assembly InstructionsSET UP Preparing the Scale for Use Replacing BatteriesOperation Instructions Description Function Only applies to pound values, not the metric values Weight Display ModeMenu Tare Weight Data ManagementAutomatic Tare Removing the TareEnter Scale Settings Operation Instructions System Settings System Test User Settings Operation Instructions Maintenance MaintenanceCleaning Disconnect the scale from the AC power sourceTroubleshooting Symptom Possible Cause Corrective ActionCalibration Path Operator Action DisplayExploded View of Scale continue Parts List Description QtyWarranty Limited WarrantyWhat does the Warranty Cover? To get Warranty Service