Omega Engineering CN8260, CN8240 Param, Error, Type character Message Type, Code Description

Page 10

CN8240 & CN8260 Series Communications & Options

Digital

Communications Option

Type

This is a single character identifying the type of message. The following table lists the type characters for all messages.

TYPE character

Message Type

R

Read Request or Read

 

Response Returning a Positive Result

r

Read Response Returning a Negative Result

W

Write Positive Value Request and Response

w

Write Negative Value Request and Response

A

Auxiliary Command

For further information, see following sections on the different message types.

PARAM

This is a two character, message specific, parameter ID. For a Read/Write Request or Response message, this ID identifies the controller parameter. For an Auxiliary Command Request or Response message, this ID specifies the command.

 

ERROR

Code

Description

 

 

0

No error.

1

Framing error.

2

Hardware error.

3

Parity error.

4

Bad character in the TYPE field.

5

Bad message. Message cannot be understood.

6

Bad checksum. The checksum received

 

did not match the checksum of the message.

7

Bad zone ID.

8

Bad auxiliary command ID. The auxiliary

 

command is not supported in this controller.

9

Bad parameter ID. The parameter is not

 

supported in this controller.

A

Bad data. Bad representation in the data field

 

or data is out of range.

B

Attempt to write to a read only parameter.

C

Cannot write to a parameter because it’s in use.

7

Image 10
Contents Series Omega.comTM Table of Contents Option -C2 RS-232 one-to-one Communications Option -AL1, -AL2 Relay, N.O. Alarm OutputLoad Type Voltage Limit Options -XP1, -XP2, -XP3, -XP4 Transducer Power Supply Option -C4 RS-485 one-to-many CommunicationsOption -DIC Options -PVSV1, -PVSV2 Process OutputOptions -DIC, -RSP4, -RSP5, -RSP6, -RSP7 Remote Setpoint Options -RSP4, -RSP5, -RSP6, -RSP7With Wiring diagram for digital communications Remote Communications OptionsStart Char Omega+ ProtocolMessage Formats ZoneParam ErrorType Type character Message TypeChksum END CharWrite Command to Controller Read Command to ControllerRead Response from Controller Write Response from Controller Auxiliary Command to ControllerAuxiliary Response from Controller Read Request Request Message DescriptionWrite Request Field Description TypeResponse Messages Auxiliary Command RequestField Description Read Response Write ResponseAuxiliary Command Response CN8240/CN8260 Parameter CodesNumber Name Read Write Data Field Value RAM, EepromDigital Communications Option PID Digital Communications Option Digital Communications Option Digital Communications Option Auxiliary Commands Or RTD w/ DecimalCommand Clear Latched Alarms Parameter # Description Command Retrieve Display Parameter # DescriptionRequest Data Fields CN8240 & CN8260 Series Communications & Options CN8240 & CN8260 Series Communications & Options CN8240 & CN8260 Series Communications & Options Warranty / Disclaimer Temperature