Emerson Process Management T-SERIES Time required to purge T-Series sensor cases

Page 17

Purge Fittings

4.Connect the supply of nitrogen or argon gas to the inlet purge connection or open inlet purge line. Leave the outlet connection open.

Exercise caution to avoid introducing dirt, moisture, rust, or other contaminants into the sensor case.

If the purge gas is heavier than air (such as argon), locate the inlet lower than the outlet, so the purge gas will displace air from bottom to top.

If the purge gas is lighter than air (such as nitrogen), locate the inlet higher than the outlet, so the purge gas will displace air from top to bottom.

5.Make sure there is a tight seal between the inlet connection and sensor case, so air cannot be drawn by suction into the case or purge line during the purging process.

6.The purge time is the amount of time required for full exchange of atmosphere to inert gas. For each sensor size, the purge time is different. Refer to Table 3. If purge lines are being used, increase the purge time to fill the additional volume of the purge line.

7.Avoid pressurizing the sensor case. At the appropriate time, shut off the gas supply, then immediately seal the purge outlet and inlet connections with the purge plugs. If pressure inside the case elevates above atmospheric pressure during operation, the flowmeter density calibration will be inaccurate.

8.Make sure the purge fitting seals are tight so air cannot be drawn by suction into the sensor case.

Table 3 Time required to purge T-Series sensor cases

 

Purge rate

Time(1)

Sensor model

ft3/hr (l/hr)

minutes

 

 

 

T025

20 (566)

1

T050

20 (566)

1

T075

20 (566)

3

T100

20 (566)

5

T150

20 (566)

10

 

 

 

(1) If purge lines are being used, increase purge time to fill the additional volume.

15

Image 17
Contents Micro Motion Series Sensor Page Troubleshooting Customer serviceSpecifications Before You BeginInstallation options European installationsDefinitions Purge plugsSeries sensor with core processor Determining a Location Hazardous area installationsFull flow tubes Environmental limitsUL and CSA ambient temperature limits Series sensor ambient and process temperature limitsMaximum cable lengths Maximum wiring distancesPipe run ValvesHygienic applications Orienting the SensorSelf-draining applications Flow direction arrow Eccentric reducerMounting a T-Series sensor Mounting the SensorCore processor to a 4-wire remote transmitter WiringWrapping the shield drain wires Applying the heat shrink Connecting the wires at the core processor Grounding Case purging procedurePurge Fittings Removing a purge plugPurge rate Time Sensor model Time required to purge T-Series sensor casesUsed equipment Return PolicyNew and unused equipment Page 20002172