Orienting the Sensor

Figure 6 Eccentric reducer

Eccentric reducer

Sensor housing

Process end connection is the same size as

 

the sensor connection.

Note: As part of the cleaning process, skid-based systems may be purged with nitrogen at the end of the cleaning cycle. When using eccentric reducers, it is possible to trap gas in the section of process piping adjacent to the reducer. Sensor performance can be impacted by intermittent flow of the captured gas in a liquid fluid stream.

The gap between the electronics housing and sensor body should be inspected periodically. Manually clean this gap when necessary.

The ASME Bioprocess Equipment Standard (ASME BPE-1997), section SC-3.12.1, includes recommendations regarding how to install process piping to ensure full drainability. In general, the ASME BPE-1997 guideline recommends a minimum drop of ¼ in/ft (2 cm/m) of piping. The angle of the process piping can have an impact on the reducer length in the sloping process line.

Flow direction arrow

The sensor has a flow direction arrow (see Figures 1–3) to help you configure the transmitter for flow direction. If possible, install the sensor so that the flow direction arrow matches actual process flow.

8

Image 10
Contents Micro Motion Series Sensor Page Specifications Customer serviceTroubleshooting Before You BeginDefinitions European installationsInstallation options Purge plugsSeries sensor with core processor Full flow tubes Hazardous area installationsDetermining a Location Environmental limitsSeries sensor ambient and process temperature limits UL and CSA ambient temperature limitsPipe run Maximum wiring distancesMaximum cable lengths ValvesSelf-draining applications Orienting the SensorHygienic applications Eccentric reducer Flow direction arrowMounting the Sensor Mounting a T-Series sensorWiring Core processor to a 4-wire remote transmitterWrapping the shield drain wires Applying the heat shrink Connecting the wires at the core processor Purge Fittings Case purging procedureGrounding Removing a purge plugTime required to purge T-Series sensor cases Purge rate Time Sensor modelNew and unused equipment Return PolicyUsed equipment Page 20002172