Ecom Instruments Ex-Pt 720 Safety regulations, Special conditions for use in Zone, Ex-Data

Page 3

-obviously incorrect measurements occur.

-measurements with the unit are no longer possible.

-the permissible limiting values were exceed.

4. Safety regulations

In order to exclude incorrect operation of the unit, its use assumes that the user is aware of and complies with the usual safety regu- lations.

-the batteries may only be changed outside the Ex-area.

-it is not allowed to take additional batteries into the Ex-area.

-only type approved batteries may be used.

-only sensors which have been approved by ecom may be used.

-It must be ensured that the unit is not taken into Zone 0.

-measurements on electrically live parts are not allowed.

-observe the measuring range of the temperature sensors.

OOverheating will damage the sensors.

-avoid use in aggressive acids or alkalis.

-the unit may not be opened in Ex-areas.

4.1 Special conditions for use in Zone 0

The measurement of temperatures with the Ex-Pt 720 within Zone 0 is only permitted if the following conditions are complied with:

-a stainless-steel sensor Type 0628 2232 is used.

-the Ex-Pt 720 itself may not be taken into Zone 0.

-It must be absolutely ensured that the unit does not find its way accidentally into Zone 0. In order to guarantee this, the carrying ststrap of the unit must be attached additionally to the belt of the user or to a loop attached to his belt. The length of the carrying ststrap must be so adjusted that when the unit hangs freely on the carrying strap, the Ex-Pt 720 maintains a spacing of 50cm to the bboundary area of Zone 0.

18

-Should the unit be used in the optionally available leather case, t˝˝the carrying strap of the unit must still comply with the conditi- ti˝ons for attachment to the user’s body.

5. Ex-Data

EC-test approval-no.:

TÜV 01 ATEX 1757 X

Ex-designation:

II 2(1) G EEx ia IIC T4

Approved for Zone 1, equipment group II, gas group C explosions- endangered gases, vapours or fog, temperature class T4.

Only the sensors specified in the approval certificate may be used. Sensor no. 0628 2232 may be used in Zone 0 !

6. Technical Details

 

 

Ambient temperature Ta:

-10 ... +50°C

Storage temperature:

-20 ... +70°C

Batteries:

9V-Block, 6LR61 as per IEC

 

(only type tested batteries)

 

 

 

Supplier

 

Type

Varta

 

Alkaline No.4822

Varta

 

Alkaline Universal No.4022

Varta

 

Alkaline Electric Power No.8022

Duracell

 

Alkaline

Duracell

 

Alkaline Ultra

Duracell

 

Professional Alkaline Battery Procell

Everady

 

 

(Ralston Energy Systems SA)

Alkaline Energizer

Panasonic

 

Alkaline Power Line Industrial Battery

Daimon

 

Alkaline

 

 

 

19

Image 3
Contents Ex-Pt Malfunctioning and inadmissible operation ContentsSafety references ApplicationTechnical Details Safety regulationsSpecial conditions for use in Zone Ex-DataHold-Function Switching-on and measuring 7.2.1 Switching-onOperating instructions CommissioningSpecial probes Switching-offBattery change Temperature probes 7.4.1 Standard probesRepairs Cleaning and maintenanceGuarantee & liability Calibration instructionsEG type approval Ex-Pt94/9/EC 89/336/EEC1st Supplement Ecom instruments GmbH