Ecom Instruments Ex-Pt 720 instruction manual Switching-off, Battery change, Special probes

Page 5

Pressing the >Hold-pushbutton< switches between °C and °F. Should no pushbutton be pressed for 5 seconds after a changeo- ver, then the new measuring value will be accepted.

7.2.5 Switching-off

Switch the unit of with the >I/O pushbutton< . The display is no longer active.

7.3 Battery change

Should the <BAT> symbol be displayed constantly, then the bat- tery will need to be changed within the next 1-2 hours. The unit switches itself automatically off should the battery voltage be too low.

Open the back of the unit by unscrewing it, remove the exhausted battery and insert a new one (observe polarity).

-while changing the battery, the cover may not be separated from t˝˝the unit top shell.

-only batteries of Type 6LR61 as per IEC may be used. The use of alalternative types of batteries is strictly forbidden.

-batteries may only be changed outside the Ex-area.

-the taking of additional batteries into the Ex-area is not allowed.

7.4Temperature probes 7.4.1 Standard probes

Immersion / insertion probe

0628 1232:

measurement range:-50 ... 400°C

 

accuracy:

class A

 

Ex-Zones:

2, 1

probe dimensions: 110mm x 4mm (w x ø)

sensor dimensions: 30mm x 3,2mm (w x ø)

response time (99%): 12 Sec.

22

Surface probe

 

 

0628 1932:

measurement range:-50 ... 400°C

 

accuracy:

class B

 

Ex-Zones:

2, 1

 

probe dimensions: 110mm x 4mm (w x ø)

 

sensor dimensions: 3,2mm (ø)

 

response time (99%):40 Sec.

Stainless-steel probe

 

0628 2232:

measurement range:-50 ... 400°C

 

accuracy:

class A

 

Ex-Zones:

2, 1, 0

 

probe dimensions: 125mm x 4mm (w x ø)

 

sensor dimensions: 15 x 3mm (w x ø)

 

response time (99%):10 Sec

7.4.2 Special probes

Immersion / insertion probe

0600

9999

measurement range:-50 ... 400°C

0628

1232:

accuracy:

class A

 

 

Ex-Zones:

2, 1

 

 

probe dimensions: >50mm x >4mm (w x ø)

 

 

sensor dimensions: >30mm x >3,2mm (w x ø)

 

 

response time (99%):12 Sec.

Surface probe

 

 

0600

9999

measurement range:-50 ... 400°C

0628

1932:

accuracy:

class B

 

 

Ex-Zones:

2, 1

 

 

probe dimensions: >50 mm x >4mm (w x ø)

 

 

sensor dimensions: >9 mm (ø)

response time (99%):40 Sec.

23

Image 5
Contents Ex-Pt Safety references ContentsApplication Malfunctioning and inadmissible operationSpecial conditions for use in Zone Safety regulationsEx-Data Technical DetailsOperating instructions Switching-on and measuring 7.2.1 Switching-onCommissioning Hold-FunctionBattery change Switching-offTemperature probes 7.4.1 Standard probes Special probesGuarantee & liability Cleaning and maintenanceCalibration instructions Repairs94/9/EC Ex-Pt89/336/EEC EG type approval1st Supplement Ecom instruments GmbH