Ecom Instruments Ex-Pt 720 instruction manual Operating instructions, Commissioning, Hold-Function

Page 4

Operating life:

approx. 100h

 

Dimensions:

190 x 57 x 42 mm

 

Weight:

approx. 200g (with batteries)

Case material:

ABS

 

 

(slight discolouration due to additives)

IP-protective rating: IP 54

 

EC-designation:

0102

 

Display:

LCD, digit height 14mm

 

Measuring device:

Pt 100-Sensor, 4-wire-technique

Measuring range:

-50°C to +400°C

 

Accuracy:

-50°C to +199,9°C:

±0,2°C, ±1 digit

 

+200°C to +400°C:

±0,2% from measured

 

 

value, ±1 digit

Resolution:

-50°C to +199,9°C:

0,1°C

 

+200°C to +400°C:

1°C

System accuracy:

-50°C to +99,9°C:

±0,6°C, ±1 digit

(with immersion/

+100°C to +400°C:

±0,5% from measured

insertion sensor)

 

 

value, ±1 digit

7. Operating instructions

Sensor connection

HOLD-button

I/O button

Safety loop for

use on Zone 0

reverse side: safety screw

7.1 Commissioning

-carefully read operating instructions.

-open back of unit with key.

-insert battery correctly (observe polarity).

-insert battery safety cover.

-close unit with safety screw.

7.2Switching-on and measuring 7.2.1 Switching-on

Insert sensor into measuring socket (take note of plug coding). Switch unit on with >I/O pushbutton<.

After switching on, a short display and function test will take place. The measuring instrument is now ready for use.

7.2.2 Hold-Function

The measured values can be retained in the display with the >Hold< pushbutton

7.2.3 Auto-Off-Function

The Auto-Off-Function switches the unit off after 10 minutes ope- rating time thus avoiding unintentional discharge of the battery.

7.2.4 Changing measuring range

The display can be switched between °C and °F.

While switching the unit on with the >I/O pushbutton< press >HOLD-pushbutton< and hold it.

After the self test is finished, the currently selected max. measu- ring value will blink.

°C

20

21

Image 4
Contents Ex-Pt Contents Safety referencesApplication Malfunctioning and inadmissible operationSafety regulations Special conditions for use in ZoneEx-Data Technical DetailsSwitching-on and measuring 7.2.1 Switching-on Operating instructionsCommissioning Hold-FunctionSwitching-off Battery changeTemperature probes 7.4.1 Standard probes Special probesCleaning and maintenance Guarantee & liabilityCalibration instructions RepairsEx-Pt 94/9/EC89/336/EEC EG type approval1st Supplement Ecom instruments GmbH