4.Taking medication

4.When treatment is complete, turn the power off. Check that there is no condensation or moisture in the air tube and unplug the device from the power outlet.

Power Outlet

Warning:

Condensation may occur in the air tube. Do not store the air tube while it contains condensation or moisture. This could result in infection via bacteria.

If there is moisture or liquid remaining in the air tube, be sure to follow the procedure below to remove any moisture from within the air tube.

1)Make sure that the air tube is still connected to the air connector on the compressor.

2)Unplug the air tube from the nebuliser kit.

3)Turn on the compressor and pump air through the air tube to expel the moisture.

18

Image 18
Contents Model NE-C30-E Contents Introduction Intended useImportant safety instructions UsageImportant safety instructions Risk of electrical shock General Safety Precautions Save these instructions for future referenceKnow your unit Rear View Bottom ViewAccessories Guide to using the Comp Air Elite Connecting the AC adapter Preparing the nebuliser for useAdding medication and assembling the parts Attaching the adult mask PVC Attaching the child mask PVC Preparing the nebuliser for use Taking medication To stop nebulisationUsing the mouthpiece Power Outlet Cleaning and daily disinfecting Disinfecting the nebulising parts Handling the Vaporiser Head Maintenance and storage Changing the air filter Troubleshooting Symptom Cause RemedyTroubleshooting Technical data Mmad = Mass Median Aerodynamic Diameter == TypeTypeBB Technical data Optional parts and accessories Filter Cover Carrying Bag Using the battery for Important notes on using the optional rechargeable battery Attach the optional battery to When battery power is low Made in China