| 8. Troubleshooting | |
|
|
| |
Symptom | Cause | Remedy | |
|
|
| |
The device is very hot. | The compressor is covered | Remove the cloth. | |
by a cloth. | |||
|
| ||
|
|
| |
| Is the optional rechargeable | Charge the battery. | |
| battery charged sufficiently? | ||
|
|
| |
| The battery has expired and |
| |
The device stops soon when | cannot be recharged | Replace with a new battery. | |
using the optional | enough. |
| |
rechargeable battery, or | Has the battery been left | Charge the battery. | |
operation time is short. | unused for a long time? | ||
| |||
|
|
| |
| Is the battery allowed to fully | Rundown, then recharge the | |
| rundown before it is | battery two or three times. | |
| recharged? | (See Section 11) | |
|
|
|
Notes:
•If the suggested remedy does not solve the problem, do not try to repair the device - no parts of the unit are user serviceable.
Return the unit to an authorized OMRON retail outlet or distributor.
•The device should be serviced once a year by an approved OMRON service centre.
EN