CDTX-90-3 Options Menu
Display |
| Description |
(Factory settings shown) |
| |
|
| |
|
|
|
Contrast: |
| Adjust the LCD contrast for best viewing. A setting of 1 is lower contrast, 5 is higher. |
3 | > | In general, select lower contrast if the display is in warmer ambient surroundings. |
| ||
|
|
|
Cond Decimal: |
| Set the decimal to the best resolution for your application. The display will |
****.* | > | automatically scale down to this restriction. Select *****., ****.*, ***.** **.*** or |
*.**** | ||
|
|
|
Averaging |
| OFF provides the most instantaneous response to process changes. Select |
Off | > | LOW (4 sec) or HIGH (8 sec) averaging if your process experiences frequent or |
extreme fluctuations. | ||
|
|
|
Loop1 Adjust: |
|
4.00 mA | > |
|
|
Loop1 Adjust: |
|
20.00 mA | > |
|
|
Adjust the minimum and maximum current output. The display value represents the precise current output. Adjustment limits:
•3.80 mA < 4.00 mA > 5.00 mA
•19.00 mA < 20.00 mA > 21.00 ma
Use this setting to match the system output to any external device.
Temp Display:
ºC >
Select ºC or ºF
Temperature |
|
Comp %: 2.00 | > |
Set the percent change in Conductivity caused by a 1ºC change in temperature. May be from 0.00 to 10.00.
Output1 Active:
Low >
Active HIGH: This setting is used to turn a device (pump, valve) ON at the setpoint.
Active LOW: This setting is used to turn a device OFF at the setpoint.
Test Loop 1:
>
Press UP and DOWN keys to manually order any output current value from 3.6 mA to 21.00 mA to test the output loop.
Test Output 1:
>
Press UP and DOWN keys to manually toggle the Open Collector output state.
Settings will repeat for Loop 2 and Output 2.
page 15