Omega CDTX-90 manual Installation, Specifications

Page 3

Omega CDTX-90 Series Conductivity/Resistivity Transmitter

CAUTION!Contents

•

Remove power to unit before wiring

1.

Installation

 

input and output connections.

2.

Specifications

•

Follow instructions carefully to avoid

3.

Electrical Connections

 

personal injury.

4.

Menu Functions

 

 

 

 

Conductivity/Resistivity

62.50uS/cm

25.0°C

ENTER

1. Installation

CDTX-90 Series transmitters are available in two styles: panel mount and field mount. The panel mount is supplied with the necessary hardware to install the transmitter. This manual includes complete panel mounting instructions.

Field mounting requires a separate mounting kit. The FP90UM Universal kit enables the transmitter to be installed virtually anywhere. Detailed instructions for field installation options are included with the FP90UM Universal kit.

1.1Panel Installation

1.The panel mount transmitter is designed for installation using a 1/4 DIN Punch. For manual panel cutout, an adhesive template is provided as an installation guide. Recommended clearance on all sides between instruments is 1 inch.

2.Place gasket on instrument, and install in panel.

3.Slide mounting bracket over back of instrument until quick-clips snap into latches on side of instrument.

4.To remove, secure instrument temporarily with tape from front or grip from rear of instrument. DO NOT RELEASE. Press quick-clips outward and remove.

96mm (3.8 in.)

FRONT VIEW

Field Mount & Panel Mount

96 mm

82 mm

(3.8 in.)

(3.23 in.)

106 mm (4.18 in.)

SIDE VIEW

Field Mount

 

 

gasket

 

 

panel

 

 

 

 

 

 

 

 

 

 

mounting

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

terminals

bracket

 

Optional

92 mm

 

 

 

 

 

 

 

 

Rear

 

 

 

 

 

 

 

 

(3.6 in.)

 

 

 

 

 

 

 

 

Cover

 

 

 

 

 

 

 

 

latch

 

 

 

 

 

 

 

 

 

 

 

 

 

 

ensr Gn)d

 

 

 

 

 

 

 

 

7

(SS HIELD

 

 

 

2

LSoyostpem-

PPwwrr

4

Output -

65

(S(SRBeeELnnADssrrC) IVKN+)

 

 

 

 

System

 

 

 

 

 

 

 

 

1

Loop +

 

3

Output +

 

 

 

 

56 mm

 

 

 

 

 

 

 

 

41 mm

(2.2 in.)

 

 

 

 

 

 

 

 

(1.6 in.)

97 mm

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

(3.8 in.)

 

 

 

 

 

 

 

 

SIDE VIEW

Panel Mount

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Panel Mount

Installation Detail

 

quick-clips

2. Specifications

General

Compatible electrodes: Omega CDCE-90 Series Standard and Certified Series Conductivity/Resistivity Electrodes

Accuracy: ±2% of reading Enclosure:

•

Rating:

NEMA 4X/IP65 front

•

Case:

PBT

•Panel case gasket: Neoprene

•

Window:

Polyurethane coated polycarbonate

•

Keypad:

Sealed 4-key silicone rubber

•

Weight:

Approx. 325g (12 oz.)

Display:

•Alphanumeric 2 x 16 LCD

•Contrast: User selected, 5 levels

•Update rate: 1.8 seconds

Electrical

•Power: 12 to 24 VDC ±10%, regulated, 80 mA max.

Sensor input range:

•Conductivity: 0.055 to 400,000 ∝S

• Resistivity: 10 KΩ to 18.2 MΩ

(Measurements from 10 MΩ to 18 MΩ (0.055 ∝S to 0.1 ∝S) must be performed in solution temperatures from 20 °C to 100 °C. )

• TDS:

0.023 to 200,000 ppm

•Temperature: PT 1000, -25 to 120°C (-13 to 248°F) Current output:

•4 to 20 mA, isolated, fully adjustable and reversible

• Max loop impedance:

50

Ω max. @ 12 V

 

325

Ω max. @ 18 V

 

600

Ω max. @ 24 V

•Update rate: 200 mS

•Accuracy: ±0.03 mA @ 25°C, 24 V Relay outputs

(CDTX-90-2 includes 2 mechanical SPDT contacts):

•Maximum voltage rating:

5 A @ 30 VDC, or 5 A @ 250 VAC, resistive load

•Programmable for:

•High or Low setpoint with adjustable hysteresis

•Pulse operation (max. rate 400 pulses/min)

Open-collector output, optically isolated:

•50 mA max. sink, 30 VDC maximum pull-up voltage.

•Programmable for:

•High or Low setpoint with adjustable hysteresis

•Pulse operation (max. rate 400 pulses/min)

Environmental

•Operating temperature: -10 to 70°C (14 to 158°F)

•

Storage temperature:

-15 to 80°C (5 to 176°F)

•

Relative humidity:

0 to 95%, non-condensing

•

Maximum altitude:

2000 m (6562 ft)

•

Insulation category:

II

•

Pollution degree:

2

Standards and Approvals

•CSA, CE, UL listed

•Immunity: EN50082-2

•Emissions: EN55011

• Safety: EN61010

•Manufactured under ISO 9001 and ISO 14001

•U.S.A. Patent # 5,708,363

page 3

Image 3
Contents User’s Guide Internet e-mail info@omega.com Specifications InstallationSensor Connections Electrical Connections Sensor SelectionCDTX-90-1 Terminal connections CDTX-90-1 System Power/Loop ConnectionsTerminals 5-10 Relay Outputs CDTX-90-2 Terminal ConnectionsLoop System Pwr Output Loop + Terminals 3-6 Loop PowerCDTX-90-3 Terminal Connections Loop Output Loop 2+ System PwrCDTX-90 Series View Menu Made an Error? Finished Editing?Step CDTX-90-1 Calibrate Menu Contrast CDTX-90-1 Options MenuCDTX-90-2 Calibrate Menu CDTX-90-2 Options Menu CDTX-90-3 Calibrate Menu CDTX-90-3 Options Menu Settings will repeat for Loop 2 and OutputRequirements Calibration ProcedureCalibration with Nist Traceable Solutions Temperature CoefficientTroubleshooting Page WARRANTY/DISCLAIMER OMEGA…Of Course