3. Electrical Connections

Sensor Selection

Select the sensor with a range closest to your process minimum and maximum values.

CDCE-90-001 (0.01 cell): 0.055 to 100 S (10 kto 18 M) Use the CDCE-90-001 for all resistivity requirements

CDCE-90-01 (0.1 cell):

1 to 1000 S

CDCE-90-1 (1.0 cell):

10 to 10,000 S

CDCE-90-10 (10.0 cell): 100 to 200,000 S

CDCE-90-20 (20.0 cell): 200 to 400,000 S

Wiring Tips:

Do not route sensor cable in conduit containing AC power wiring. Electrical noise may interfere with sensor signal.

Routing sensor cable in grounded metal conduit will help prevent electrical noise and mechanical damage.

Seal cable entry points to prevent moisture damage.

Only one wire should be inserted into a terminal. Splice double wires outside the terminal.

Sensor cable may not be extended over 100 ft.

Ultra-

 

 

 

 

 

 

 

 

 

 

CDCE-90 Series Sensor Ranges

 

Pure

Pure

Rinse

 

 

 

 

 

 

 

 

 

 

 

Models

 

 

 

 

 

 

 

 

 

 

 

 

CDCE-90-20

 

 

 

 

 

 

 

 

 

 

CDCE-90-1

 

 

 

 

 

 

 

 

 

 

 

 

 

 

CDCE-90-10

 

 

 

Sensor

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

CDCE-90-01

 

 

 

 

 

 

CDCE-90-001

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

0.055

1

10

100

200

 

 

1,000

10,000

200,000

400,000

(18 MΩ)

 

 

 

 

 

 

(10 kΩ)

 

 

 

 

Conductivity Range (µS)

CDTX-90

Conductivity/Resistivity

≤ 100 ft. (30 m)

25.0°C

62.50 uS/cm

 

ENTER

 

Caution: Failure to fully open terminal jaws before removing wire may permanently damage instrument.

Wiring Procedure

1. Remove 0.5 - 0.625 in. (13-16 mm) of insulation from wire end.

2.Press the orange terminal lever downward with a small screwdriver to open terminal jaws.

3.Insert exposed (non-insulated) wire end in terminal hole until it bottoms out.

4.Release orange terminal lever to secure wire in place. Gently pull on each wire to ensure a good connection.

Wiring Removal Procedure

1.Press the orange terminal lever downward with a small screwdriver to open terminal jaws.

2.When fully open, remove wire from terminal.

2

1

3.1 Sensor Connections

CDTX-90-1 Terminals

10

Shield (Sensor Gnd)

Black (Iso Gnd)

9

White (Temp In)

8

Red (Signal In)

7

 

CDTX-90-2 Terminals

14

Shield (Sensor Gnd)

Black (Iso Gnd)

13

White (Temp In)

12

11

Red (Signal In)

 

CDCE-90 Series Standard and Certified Cells

CDTX-90-3 Terminals

14

Shield (Sensor Gnd)

Black (Iso Gnd)

13

White (Temp In)

12

11

Red (Signal In)

page 4

Image 4
Contents User’s Guide Internet e-mail info@omega.com Installation SpecificationsElectrical Connections Sensor Selection Sensor ConnectionsCDTX-90-1 System Power/Loop Connections CDTX-90-1 Terminal connectionsCDTX-90-2 Terminal Connections Terminals 5-10 Relay OutputsTerminals 3-6 Loop Power CDTX-90-3 Terminal ConnectionsLoop Output Loop 2+ System Pwr Loop System Pwr Output Loop +CDTX-90 Series View Menu Finished Editing? Made an Error?Step CDTX-90-1 Calibrate Menu CDTX-90-1 Options Menu ContrastCDTX-90-2 Calibrate Menu CDTX-90-2 Options Menu CDTX-90-3 Calibrate Menu Settings will repeat for Loop 2 and Output CDTX-90-3 Options MenuCalibration Procedure Calibration with Nist Traceable SolutionsTemperature Coefficient RequirementsTroubleshooting Page WARRANTY/DISCLAIMER OMEGA…Of Course