JDS Uniphase RM3750 user manual Rear Panel, Termination Techniques, Component Function

Page 23

RM3 SERIES BACKREFLECTION METER

USER’S MANUAL

Rear Panel

The back of the meter (Figure 4) can vary with the model.

-

25 VA max Fuse 1A/250V

Figure 4: Back of the Meter

The components on the rear panel are described in Table 6.

Table 6: Rear Panel Components

Component

Function

 

 

CAL

Access to the User CAL value, for calibration purposes.

 

 

LOG

Foot pedal jack for data logging.

 

 

RS232C

RS232 serial interface port.

 

 

Termination Techniques

Termination is a technique used to block all backreflections beyond a certain point in a network of components. In all measurement testing, two terminations must be made: the first (for measuring BR0) is made before the DUT, for example, on the RM meter side; the second (for measuring BRtot) is made after the DUT. Both termination points need to be made as close as possible to the DUT to minimize the errors associated with Rayleigh backscattering in the fiber. The backreflection of all components and connections between these two points is included in the equation for calculating the backreflection value (BR) of the DUTs.

Termination Techniques for Single-mode Fiber (except 980 nm)

Termination for single-mode (except 980nm) is performed by mandrel wrapping.

High-attenuation bends in the fiber (bends with a relatively small radius) remove all backreflection and can be made anywhere along the length of the cable. When measuring backreflection levels below -60 dB, low-attenuation bends (bends with a relatively large radius)

10112341 Rev 002

Page 23 of 39

Image 23
Contents User’s Manual RM3 Series Backreflection MeterContents Programming Guide Customized Features and Test DataSafety Information Safety Information, Instructions, and SymbolsClass Laser Product Operating the Unit Safety InstructionsBefore Initializing and Operating the Unit RM3 Series Backreflection Meter USER’S Manual Proceeding Safety SymbolsSymbol FDA-CDRH Compliance CSA / IEC Compliance ComplianceCE Compliance RM3 Series Backreflection Meter USER’S Manual General Information General Information and SpecificationsBackreflection Measurements Loss and Power MeasurementsHybrid Jumpers Key FeaturesOutput Port Internal Light SourcesApplications Standard AccessoriesOptional Accessories Parameter Specification Single-mode Multimode SpecificationsFlexcor is a trademark of CorningR Environmental ElectricalPhysical Initial Inspection Getting StartedBefore Initializing and Operating the Unit Storing and Shipping Operating Environment Cleaning Connectors Connector Cleaning connector type can vary Powering Up the MeterKey/LED Description Operating and Maintenance InstructionsFront Panel Termination Techniques for Single-mode Fiber except 980 nm Rear PanelTermination Techniques Component FunctionMeasuring Backreflection for Single-mode except 980 nm Backreflection Measurements Setting Up the Meter for Backreflection MeasurementsMeasuring Backreflection Backreflection Accuracy and Range Loss and Power MeasurementsSetting Up the Meter for Loss and Power Measurements Loss and Power Measurements Measuring PowerMeasuring Absolute Power Measuring Relative Power Insertion LossDisplay Description Data Logging Using the Foot PedalMessages and Symbols Power AccuracyStabilizing Calibrating the MeterPower Calibration Adjustment Backreflection Calibration AdjustmentMaintaining the Meter Cleaning the UnitCleaning the Connector Ends Troubleshooting Recharging the BatteryReplacement of Internal Fuses Reflections Before the DUT Front Panel ConnectorsLong Cables for single mode model only, except 980 nm model Loss Before the DUTUp the Meter for Backreflection Other Display Problem SolutionLaser Stability CalibrationRS232 Serial Interface RS232 Interface SpecificationsName Symbol Pin Number Signal Direction Programming GuideCommand Action RS232-GPIB Converter Left Dip SwitchSwitch Setting Meaning Right Dip SwitchRS232 Interface Program Example Programming ExamplesCustomized Features and Test Data