Fluke 787 user manual Display Element MA, DC, mV, AC Or kΩ, kHz Auto 400100030

Page 20

787

Users Manual

Table 6. Display (cont.)

No.

J

K

L

Element

mA, DC, mV, AC,

M or kΩ, kHz

Auto

400100030

EPN

Meaning

Show the input or output units and multipliers associated with the numerals

Range status indicators:

Auto means autoranging is on.

The number plus the unit and multiplier indicate the active range.

One of these lights in mA ramping or step output (rotary switch position mA J):

Emeans slow continuous 0% - 100% - 0% ramping.

Pmeans fast continuous 0% - 100% - 0% ramping.

Nmeans ramping in 25% steps.

16

Image 20
Contents 787 Limited Warranty & Limitation of Liability Table of Contents 787 IndexContacting Fluke Read Safety Information before you use the meterIntroduction Safety Information 787ProcessMeter International Symbols Meaning Symbol MeaningHow to Get Started Fluke 787 ProcessMeterGetting Acquainted with the Meter Input/Output JacksSimulate Transmitter Function Input/Output Jacks Measurement FunctionsSource Current Function Rotary Switch Positions for Measurements Functions Pushbutton ActionsRotary Switch Positions for mA Output Default Function Pushbuttons Pushbuttons Functions Pushbutton CoarseElements of the Display Element Display MeaningDisplay Element MA, DC, mV, AC Or kΩ, kHz Auto 400100030 Measuring a Composite Signal Measuring Electrical ParametersInput Impedance RangesTesting Diodes Displaying Minimum, Maximum, and AverageUsing TouchHold Compensating for Test Lead ResistanceUsing the Current Output Functions Source ModeSourcing Current Simulate Mode Changing the Current SpanSimulating a Transmitter Producing a Steady mA Output MA Output Adjust PushbuttonsAdjustment Manually Stepping the mA Output MA Stepping PushbuttonsMA Step Values Value for each span setting To 20 mA Auto Ramping the mA OutputPower-Up Options Power-Up OptionsPushbutton Default Action TakenBattery Life Using the Holster and Flex-StandMaintenance General MaintenanceUsing the Holster and Flex-Stand Replacing the Battery Replacing the BatteryReplacing a Fuse If the Meter does not Work Replacing a FuseReplacement Parts and Accessories Replacement Parts DescriptionFluke PN or Model no QuantityMP8 Specifications ResolutionDC Volts Measurement Range V dcAC Volts Measurement Range mV dc Resolution 400.0 Accuracy % of Reading + CountsDC Millivolts Measurement AC Current Measurement DC Current MeasurementMeasurement Current Range ResolutionOhms Measurement Input Range Minimum Sensitivity rms Sinewave 5 Hz to 5 kHz Frequency Counter AccuracyFrequency Counter Sensitivity General Specifications Source modeSimulate Mode Diode Test and Continuity TestCertifications CSA, UL, TÜV Index Knob positions, 9 Pushbuttons Loop supply, external Index 787