Fluke 787 user manual Specifications, DC Volts Measurement, Range V dc, Resolution

Page 40

787

Users Manual

Specifications

All specifications apply from +18°C to +28°C unless stated otherwise.

All specifications assume a 5 minute warmup period.

The standard specification interval is 1 year.

Note

“Counts” means number of increments or decrements of the least significant digit.

DC Volts Measurement

Range (V dc)

4.000

40.00

400.0

1000

Resolution

0.001V

0.01V

0.1V

1V

Accuracy, ±(% of Reading + Counts)

0.1% + 1

0.1% + 1

0.1% + 1

0.1% + 1

Input impedance: 10 MΩ (nominal), < 100 pF

Normal mode rejection ratio: >60 dB at 50 Hz or 60 Hz

Common mode rejection ratio: >120 dB at dc, 50 Hz, or 60 Hz

Overvoltage protection: 1000V

36

Image 40
Contents 787 Limited Warranty & Limitation of Liability Table of Contents 787 IndexIntroduction Read Safety Information before you use the meterContacting Fluke Safety Information 787ProcessMeter International Symbols Meaning Symbol MeaningHow to Get Started Fluke 787 ProcessMeterGetting Acquainted with the Meter Input/Output JacksSource Current Function Input/Output Jacks Measurement FunctionsSimulate Transmitter Function Rotary Switch Positions for Measurements Functions Pushbutton ActionsRotary Switch Positions for mA Output Default Function Pushbuttons Pushbuttons FunctionsPushbutton CoarseElements of the Display Element Display MeaningDisplay Element MA, DC, mV, AC Or kΩ, kHz Auto 400100030 Measuring a Composite Signal Measuring Electrical ParametersInput Impedance RangesTesting Diodes Displaying Minimum, Maximum, and AverageUsing TouchHold Compensating for Test Lead ResistanceUsing the Current Output Functions Source ModeSourcing Current Simulate Mode Changing the Current SpanSimulating a Transmitter Adjustment MA Output Adjust PushbuttonsProducing a Steady mA Output Manually Stepping the mA Output MA Stepping PushbuttonsMA Step Values Value for each span setting To 20 mA Auto Ramping the mA OutputPower-Up Options Power-Up OptionsPushbutton Default Action TakenBattery Life Using the Holster and Flex-StandMaintenance General MaintenanceUsing the Holster and Flex-Stand Replacing the Battery Replacing the BatteryReplacing a Fuse If the Meter does not Work Replacing a FuseReplacement Parts and Accessories Replacement Parts DescriptionFluke PN or Model no QuantityMP8 Specifications ResolutionDC Volts Measurement Range V dcDC Millivolts Measurement Range mV dc Resolution 400.0 Accuracy % of Reading + CountsAC Volts Measurement AC Current Measurement DC Current MeasurementOhms Measurement Range ResolutionMeasurement Current Frequency Counter Sensitivity Frequency Counter AccuracyInput Range Minimum Sensitivity rms Sinewave 5 Hz to 5 kHz General Specifications Source modeSimulate Mode Diode Test and Continuity TestCertifications CSA, UL, TÜV Index Knob positions, 9 Pushbuttons Loop supply, external Index 787