IDEAL INDUSTRIES 61-095 technical manual Measurement Characteristics

Page 6

Page 4

Measurement Characteristics

Accuracy is ±(% reading + number of digits) at 23ºC ± 5ºC, less than 80% R.H.

Voltage

Function

Range

Accuracy

Overload protection

 

 

 

 

V

600 Vrms

± (1.5%+ 3 dgt) 40Hz ~ 500Hz

600 Vrms

 

 

 

 

V

600 V

± (1% + 2 dgt)

 

 

 

 

 

Input impedance: 1MΩ // less than 100pF.

Resistance & Continuity

Function

Range

Accuracy

Overload protection

 

 

 

 

2000 Ω

± (1% + 2 dgt)

600 Vrms

 

 

 

 

Max. open circuit voltage: 3V

Continuity check: Internal sounds activates if the resistance of the circuit under test is less than 25 Ω. It will then turn off if the resistance is increased beyond 400 Ω.

AC Current

Function

Range

Accuracy

Overload protection

 

 

 

 

A

200.0A

± (3% + 3 dgt)*45Hz ~ 66Hz

200A

 

 

 

 

* Adjacent conductor influence: <0.05 A/A

Form number TM61095

Rev 2 September 2002

Image 6
Contents Ideal INDUSTRIES, INC Technical Manual Table of Contents Precautions and Safety Information Symbols Table A. The Symbols MeaningSafety Specific precautionsSpecifications General specifications Characteristics DescriptionMeasurement Characteristics Voltage Function Range Accuracy Overload protectionEnvironmental Characteristics Description Certifications and CompliancesRequired Equipment Cleaning Basic MaintenanceReplacing the Battery Opening the Meter CaseAC Voltage Test Step Input Frequency Reading Testing the Voltage FunctionPerformance Tests DC Voltage Test Step Input ReadingAmp a Function Testing the Resistance FunctionResistance Test Step Source Reading ACA Test Step Source Frequency ReadingTop shows correct method bottom shows incorrect method ACV Calibration Adjust VR2 See FigureACA Calibration Adjust VR3 See Figure CalibrationCalibration Adjustment Points