IDEAL INDUSTRIES 61-095 technical manual Required Equipment

Page 8

Page 6

Required Equipment

Required equipment is listed in Table B. If the recommended models are not available, equipment with equivalent specifications may be used.

Repairs or servicing should be performed only by qualified personnel.

Table B. Required Equipment

Equipment

Required Characteristics

Recommended Model

 

 

 

 

AC Voltage Range: 0-750V AC

 

 

Accuracy: ±0.07% (Basic)

 

 

Frequency Range: 40 ~ 500Hz

Fluke 5500 or Wavetek

 

Accuracy: ± 2%

 

9100 Calibrator or

 

 

 

DC Voltage Range: 0-1000V DC

Equivalent

 

 

Calibrator

Accuracy: ±0.006% (Basic)

 

 

Current Range: 0 ~ 100A

 

 

Accuracy: AC (40Hz to 100Hz): ±0.08% (Basic)

Ballantine 1620A

 

 

 

Resistance Range: 1Ω ~ 2000Ω

 

 

Accuracy: ±0.03% (Basic)

 

 

 

 

Form number TM61095

Rev 2 September 2002

Image 8
Contents Ideal INDUSTRIES, INC Technical Manual Table of Contents Precautions and Safety Information Symbols Table A. The Symbols MeaningSafety Specific precautionsSpecifications General specifications Characteristics DescriptionMeasurement Characteristics Voltage Function Range Accuracy Overload protectionEnvironmental Characteristics Description Certifications and CompliancesRequired Equipment Basic Maintenance Replacing the BatteryCleaning Opening the Meter CaseTesting the Voltage Function Performance TestsAC Voltage Test Step Input Frequency Reading DC Voltage Test Step Input ReadingTesting the Resistance Function Resistance Test Step Source ReadingAmp a Function ACA Test Step Source Frequency ReadingACV Calibration Adjust VR2 See Figure ACA Calibration Adjust VR3 See FigureTop shows correct method bottom shows incorrect method CalibrationCalibration Adjustment Points