Hach 3650 manual Configuring the Instrument, Automatic Data Acquisition Setting Sampling Intervals

Page 32

26 of 50

Options Setup - 3650 Portable O2 Analyzer

 

 

3.4 Configuring the Instrument

The ORBISPHERE 3650 analyzer can be readily configured for your application using the following commands in the Configuration menu. The instrument must be connected to your PC in order to change its configuration.

3.4.1 Automatic Data Acquisition - Setting Sampling Intervals

The analyzer can perform as a standalone data acquisition device, automatically recording gas measurements with the date and time, and storing up to 500 of these values. Choosing the WinLog97 program's Configuration, Sampling Rate menu lets you select time intervals (acquisition rate) for this storage capability.

Use the slide bar to view and select a sampling rate, from 15 seconds to 1 hour. The selected rate is shown in the Acquisition rate window.

Click OK to save this rate. Once your choice is made, the analyzer can be used independently of the WinLog97 program for data acquisition, as described in “Automatic Data Acquisition” on page 16.

Fig 3-9: Automatic Data Acquisition

Note:

The Acquisition rate set via this menu is independent from the monitoring chart updating rate described in “Monitoring Measurements In Real-Time” on page 21. The sampling rate menu applies only to automatic data acquisition, while the chart updating rate is used only for displaying real-time results via the monitoring chart.

3.4.2 Membrane Selection

You may find it necessary to use a different type of membrane for different applications. Naturally, with any membrane change, you will need to re-calibrate (see “Sensor Calibration” on page 32). You should also consider the changes in required flow rates and response times, which are specified in the accompanying EC Sensors - Maintenance & Installation Manual.

To re-configure the analyzer, choose Configuration, Membrane to bring up the box which reveals the membrane models available.

Choose OK when the desired membrane is selected.

Fig 3-10: Select Membrane

Note:

Only those membranes applicable for your instrument model can be selected. All other membranes are grayed out.

ORBISPHERE

Operator Manual

Image 32 Contents
Operator Manual Page Product Recycling Information Orbisphere Product Disposal Following only applies to European customersRestriction of Hazardous Substances MarkingTable of Contents Accessories and Attachments Maintenance and Troubleshooting SpecificationsCalibrations Part ListsSafety Conventions Safety Precautions Manual OverviewAbout this Manual Revision HistorySafety Recommendations Service and RepairsPrecautionary Labels AcknowledgementsOrbisphere Installation Sensor Installation Flow Chamber InstallationWinLog97 PC Program Installation External Power optionalSample Tube Adapter optional ConnectionsInstrument PC Connection Installation Completion Check List Instrument Clock SettingBatteries Electrochemical SensorStorage When not Used Flow ChamberOperating Information Operating ControlsOperating Information 3650 Portable O2 Analyzer Taking Measurements Preconditioning SensorStoring Measurements In the Instrument Automatic Data AcquisitionManual Data Acquisition Downloading Stored Values Storing and Accessing Measurements From the PCViewing Stored Measurements Altering the Sampling Point Descriptions Copying ValuesClearing Stored Values Saving ValuesPrinting Values Monitoring Measurements In Real-Time Timebase Updating Rate Maximum Samples 10 divisionsOrbisphere Options Setup Main Menu BasicsWinlog97 Export Menu Reviewing Instrument Configuration O2 Analyzer PC ConnectionMembrane Selection Configuring the InstrumentAutomatic Data Acquisition Setting Sampling Intervals Selecting Type of Calibration Select from the PCEntering a Span Gas Value Locking Out the Instrument’s CAL ButtonSensor Calibration Range Checking Dual Use model 3650/113 only Change from the PCOrbisphere Calibrations Pressure CalibrationSensor Calibration Calibration In a Span GasCalibration In Line Calibration Range Checking Calibration In AirAccessories and Attachments Accessories and Attachments 3650 Portable O2 Analyzer Maintenance and Troubleshooting MaintenanceInstrument Batteries SensorKeyboard Test TroubleshootingSerial Test Clock Settings Display TestAnalog Voltages View Measurements ViewSpecifications System SpecificationsGeneral Technical Data Instrument Display OptionsTheory of Operation Instrument AccessoriesPart Lists Spare Parts Terms and Definitions Appendix a GlossaryCommon Units Glossary 3650 Portable O2 Analyzer Annex Tables and IllustrationsAnnex 3650 Portable O2 Analyzer User Notes