26 of 50 | Options Setup - 3650 Portable O2 Analyzer |
|
|
3.4 Configuring the Instrument
The ORBISPHERE 3650 analyzer can be readily configured for your application using the following commands in the Configuration menu. The instrument must be connected to your PC in order to change its configuration.
3.4.1 Automatic Data Acquisition - Setting Sampling Intervals
The analyzer can perform as a standalone data acquisition device, automatically recording gas measurements with the date and time, and storing up to 500 of these values. Choosing the WinLog97 program's Configuration, Sampling Rate menu lets you select time intervals (acquisition rate) for this storage capability.
Use the slide bar to view and select a sampling rate, from 15 seconds to 1 hour. The selected rate is shown in the Acquisition rate window.
Click OK to save this rate. Once your choice is made, the analyzer can be used independently of the WinLog97 program for data acquisition, as described in “Automatic Data Acquisition” on page 16.
Fig
Note:
The Acquisition rate set via this menu is independent from the monitoring chart updating rate described in “Monitoring Measurements In
3.4.2 Membrane Selection
You may find it necessary to use a different type of membrane for different applications. Naturally, with any membrane change, you will need to
To
Choose OK when the desired membrane is selected.
Fig
Note:
Only those membranes applicable for your instrument model can be selected. All other membranes are grayed out.
ORBISPHERE | Operator Manual |