Agilent Technologies 6812B, 6813B, 6811B manual Examples of Front Panel Programming

Page 47

Front Panel Operation - 4

Examples of Front Panel Programming

You will find these examples on the following pages:

1Setting the output voltage amplitude

2Setting the output frequency

3Setting the dc offset

4Setting a protection feature

5Clearing a protection feature

6Generating step, pulse, and list transients

7Programming trigger delays and phase synchronization

8Programming slew rates

9Measuring peak inrush current

10Setting the GPIB address or RS-232 parameters

11Saving and recalling operating states

The examples in the ac source Programming Guide are similar to the ones in this section, except that they use SCPI commands.

1 - Setting the Output Voltage Amplitude

NOTE: The maximum voltage that the ac source can output is limited by the peak value of the waveform, which is 425 Vpeak. Since the output is programmed in units of rms volts, the maximum value that can be programmed is dependent on the peak-to-rms ratio of the selected waveform. For a sinewave, the maximum ac voltage that can be programmed is 300 Vrms. For other waveforms the maximum may be different.

When you turn on the ac source, the default output shape is a 60 Hz sinewave at 0 Vrms. There is no output from the ac source because the default output state is OFF, as indicated by the Dis annunciator.

Set the output to 120 V rms as follows:

Action

You can set the voltage in any of three ways:

1.On the Function keypad, press Voltage. On the Entry keypad, press 1 2 0 Enter. This is the easiest way to enter an accurate value.

Display

VOLT 120

2.

On the Function keypad, press Voltage. On the Entry keypad, press or ° to

VOLT

127

 

increment or decrement the existing value. This technique is useful when you are

 

 

 

making minor changes to an existing value.

 

 

3

Rotate the front panel Voltage knob to obtain 120 V. This method is best when you

120 V

60 Hz

 

want to enter a value without using the voltage menu.

 

 

 

Note: You will not see the new voltage on the front panel meter unless the output is

 

 

 

enabled.

 

 

To enable the output:

 

 

4.

On the Function keypad, press Output On/Off. The Dis annunciator will go off,

120 V

60 Hz

 

indicating that the voltage is now applied to the output terminals.

 

 

47

Image 47
Contents User’s Guide AC Power Solutions Agilent Models 6811B, 6812B, and 6813BCertification Warranty InformationGeneral Safety SummarySafety Symbols EMC DeclarationPrinting History Acoustic Noise InformationTable of Contents Performing the Verification Tests Performing the Calibration ProcedureEntry Keys Examples of Front Panel Programming IntroductionLine Voltage Conversion Error Number ListPage Location Document OrientationTopic Options Safety ConsiderationsOptions and Parts Earlier AC Source ModelsAgilent Part Number DescriptionOperator Replaceable Parts Model Description Front Panel/Remote OperationCapabilities Steady-state Output Characteristic in real-time mode Steady-state Output CharacteristicTypical Peak Current Output Capacities Peak Current/Dynamic Power CapabilityPeak Current Limit Peak Inrush ExampleCapacitance in μF 127 254 Ipeak settingReal Time Regulation RMS Current Limit CircuitVoltage Regulation Output ImpedanceOutput Coupling Page Items Supplied CleaningInspection Damage Packaging MaterialNot block the fan exhaust at the rear of the unit LocationBench Operation Rack MountingInput Source and Line Fuse Installing the Power CordInput Connections Output Connections Connecting the Power CordCurrent Ratings Wire ConsiderationsRemote Sense Connections Voltage DropsRemote Sense Connections Remote Sensing and OVP Considerations Trigger ConnectionsDigital Connections Controller Connections Gpib ConnectorRS-232 Data Format Interface CommandsRS-232 Connector RS-232 InterfaceNull Modem Interface Lines Response Data Terminator Hardware HandshakePreliminary Checkout IntroductionCheckout Procedure Using the KeypadPress 1, 6, 0, Enter Press 1, 2, 0, EnterPress Output On/Off Press ProtectCase of Trouble Error MessagesSelftest Errors Power-On Selftest ErrorsMay appear at runtime Runtime Error MessagesLine Fuse IndicatesPage Front Panel, Overall View Front Panel DescriptionAC+DC System Keys Dis lay Command FunctionScrolling Keys Function KeysImmediate Action Keys Display Display Command FunctionMeter Display Keys Display MeasurementList Output Control KeysRST CLS Protection and Status Control KeysProtclear Abort Trigger and List Control KeysInitimmed IMMThrough Are used for entering numeric values Is the decimal Entry KeysSet the output to 120 V rms as follows Setting the Output Voltage AmplitudeExamples of Front Panel Programming Action Setting the Output FrequencySetting the DC Offset To verify the output, use the meter menuClearing Protection Conditions Setting a Protection FeatureOvercurrent protection feature as follows Bit Number Bit WeightPulse Using Transient Voltage ModesVoltm Step Step TransientPulse Transient VolttDcycle List TransientVoltm Pulse WidthVolt EOL Step Auto Trigger Delays and Phase SynchronizationVoltm Fixed List Count Dwel Dwel EOL VoltSyncphas Initimmed Syncsour PhaseExample Display Programming Slew Rates Using Slew Rates to Generate WaveformsSlewt Voltm FixedSlew Step SlewmCurrpeak Syncsour Phase SyncphasMeasuring Peak Inrush Current CurrlevAction Display To set the Gpib address, proceed as follows Setting the Gpib Address and RS-232 ParametersTo configure the RS-232 interface, proceed as follows Saving and Recalling Operating StatesTable A-1. Performance Specifications1 SpecificationsTable A-1. Performance Specifications Table A-2. Supplemental Characteristics Supplemental CharacteristicsTrig Out HC TTL output Trig In 10k pullup Table A-3. Operation Below 45 Hz Operation Below 45 HzPage Equipment Required Verification and CalibrationRecommended Model Test SetupTable B-1. Equipment Required Action Normal Result Turn-On Checkout ProcedurePerforming the Verification Tests AC Voltage Programming and Measurement AccuracyVolt 50, Currrange LOW, Currlev DC Voltage Programming and Measurement AccuracyRMS Current Accuracy Test Volt 0, Outpcoup DC, OffsetRMS Current Measurement Accuracy Table B-2. Agilent 6811B Verification Test RecordTable B-3. Agilent 6812B Verification Test Record DC Programming and Measurement AccuracyTable B-4. Agilent 6813B Verification Test Record Table B-3Front Panel Calibration Menu Performing the Calibration ProcedureCalibrating and Entering Voltage Offset Values Enable Calibration ModeFront Panel Calibration Callev P4 Calibrating and Entering DC Voltage Gain ValuesCalibrating and Entering AC rms Voltage Gain Values Calibrating and Entering rms Current Values Calibrating the OVP Trip PointOFF Calibrating and Entering rms Current Measurement ValuesCalibrating the Output Impedance Saving the Calibration ConstantsCalibration Over the Gpib Changing the Calibration PasswordCalibration Error Messages Table B-5. Gpib Calibration Error MessagesFigure B-2. Calibration Program Listing Sheet 1 Figure B-2. Calibration Program Listing Sheet 2 Figure B-2. Calibration Program Listing Sheet 3 Page Table C-1. Error Numbers Error Number ListError Messages Error Messages C Page Check the Line Jumpers all Models Check the Jumper Wire Model Agilent 6811B/6812B onlyCheck the Power Transformer Connector all Models Open the UnitFigure D-1. Line Voltage Conversion Components Close the UnitIndex Index Index Japan United States Latin AmericaCanada Australia/New Zealand Europe Asia PacificManual Updates
Related manuals
Manual 187 pages 38.86 Kb Manual 20 pages 27.01 Kb Manual 3 pages 27.44 Kb

6812B, 6811B, 6813B specifications

Agilent Technologies, a leader in electronic test and measurement equipment, offers a range of powerful signal sources including the 6843A, 6834B, 6814B, 6813B, and 6811B models. These instruments are designed to support various applications in research, development, and manufacturing, providing precise signal generation capabilities.

The Agilent 6843A is a versatile signal generator known for its exceptional frequency range and modulation capabilities. It supports an extensive bandwidth, making it ideal for applications that require high-frequency signal generation. With its superior phase noise performance, the 6843A is an excellent choice for radar, wireless communications, and electronic warfare applications. The instrument features an intuitive user interface, allowing engineers to set parameters quickly and efficiently.

Next, the Agilent 6834B offers exceptional performance characteristics, including high output power and low distortion. This signal generator is particularly noted for its ability to produce complex modulation formats, making it suitable for testing advanced wireless communication systems. With a reliable and stable output, the 6834B ensures accurate and repeatable measurements, which is vital for thorough testing processes.

The 6814B model stands out for its dual-channel capabilities, allowing users to generate simultaneous signals for testing multiple components or systems. This feature significantly enhances testing efficiency and flexibility for engineers. With built-in arbitrary waveform functionality, users can create custom waveforms, making the 6814B suitable for a wide range of applications including device characterization and signal processing research.

For those seeking a more compact solution, the Agilent 6813B provides essential signal generation features without compromising on performance. It is designed for a variety of applications across telecommunications and consumer electronics, featuring a straightforward interface and robust performance metrics.

Lastly, the 6811B is an entry-level yet capable model that supports a broad spectrum of testing needs. Perfect for educational and laboratory environments, it provides essential functionalities required for effective signal generation and analysis.

Overall, Agilent Technologies' 6843A, 6834B, 6814B, 6813B, and 6811B signal generators offer an array of features and technologies that cater to various application needs. Their precision, reliability, and user-oriented designs position them as invaluable assets in any testing environment, ensuring engineers can carry out their work with confidence and accuracy.