Agilent Technologies 6813B, 6811B, 6812B manual Acoustic Noise Information, Printing History

Page 6

Acoustic Noise Information

Herstellerbescheinigung

Diese Information steht im Zusammenhang mit den Anforderungen der

Maschinenläminformationsverordnung vom 18 Januar 1991.

*Schalldruckpegel Lp <70 dB(A)

*Am Arbeitsplatz

*Normaler Betrieb

*Nach EN 27779 (Typprüfung).

Manufacturer’s Declaration

This statement is provided to comply with the requirements of the German Sound Emission Directive, from 18 January 1991.

*Sound Pressure Lp <70 dB(A)

*At Operator Position

*Normal Operation

*According to EN 27779 (Type Test).

Printing History

The edition and current revision of this manual are indicated below. Reprints of this manual containing minor corrections and updates may have the same printing date. Revised editions are identified by a new printing date. A revised edition incorporates all new or corrected material since the previous printing date.

Changes to the manual occurring between revisions are covered by change sheets shipped with the manual. In some cases, the manual change applies only to specific instruments. Instructions provided on the change sheet will indicate if a particular change applies only to certain instruments.

This document contains proprietary information protected by copyright. All rights are reserved. No part of this document may be photocopied, reproduced, or translated into another language without the prior consent of Agilent Technologies. The information contained in this document is subject to change without notice.

© Copyright 1995, 1997, 1998 Agilent Technologies, Inc.

Edition 1 _________August, 1995

 

Edition 2

_________February, 1997

 

Edition 3

_________March, 1998

 

Edition 4

_________December 1998

 

Update

_________April 2000

6

Image 6
Contents AC Power Solutions Agilent Models 6811B, 6812B, and 6813B User’s GuideWarranty Information CertificationSafety Summary GeneralSafety Symbols Declaration EMCAcoustic Noise Information Printing HistoryTable of Contents Introduction Performing the Calibration ProcedureEntry Keys Examples of Front Panel Programming Performing the Verification TestsError Number List Line Voltage ConversionPage Document Orientation TopicLocation Earlier AC Source Models Safety ConsiderationsOptions and Parts OptionsDescription Operator Replaceable PartsAgilent Part Number Front Panel/Remote Operation CapabilitiesModel Description Steady-state Output Characteristic Steady-state Output Characteristic in real-time modePeak Inrush Example Peak Current/Dynamic Power CapabilityPeak Current Limit Typical Peak Current Output CapacitiesIpeak setting Capacitance in μF 127 254Output Impedance RMS Current Limit CircuitVoltage Regulation Real Time RegulationOutput Coupling Page Damage Packaging Material CleaningInspection Items SuppliedRack Mounting LocationBench Operation Not block the fan exhaust at the rear of the unitInstalling the Power Cord Input ConnectionsInput Source and Line Fuse Connecting the Power Cord Output ConnectionsVoltage Drops Wire ConsiderationsRemote Sense Connections Current RatingsRemote Sense Connections Trigger Connections Digital ConnectionsRemote Sensing and OVP Considerations Gpib Connector Controller ConnectionsRS-232 Interface Interface CommandsRS-232 Connector RS-232 Data FormatHardware Handshake Null Modem Interface Lines Response Data TerminatorIntroduction Preliminary CheckoutUsing the Keypad Checkout ProcedurePress Protect Press 1, 2, 0, EnterPress Output On/Off Press 1, 6, 0, EnterPower-On Selftest Errors Error MessagesSelftest Errors Case of TroubleIndicates Runtime Error MessagesLine Fuse May appear at runtimePage Front Panel Description Front Panel, Overall ViewAC+DC Dis lay Command Function System KeysFunction Keys Immediate Action KeysScrolling Keys Display Measurement Display Command FunctionMeter Display Keys DisplayOutput Control Keys ListRST Protection and Status Control Keys ProtclearCLS IMM Trigger and List Control KeysInitimmed AbortEntry Keys Through Are used for entering numeric values Is the decimalSetting the Output Voltage Amplitude Examples of Front Panel ProgrammingSet the output to 120 V rms as follows To verify the output, use the meter menu Setting the Output FrequencySetting the DC Offset ActionBit Number Bit Weight Setting a Protection FeatureOvercurrent protection feature as follows Clearing Protection ConditionsUsing Transient Voltage Modes PulseVoltt Step TransientPulse Transient Voltm StepWidth List TransientVoltm Pulse DcycleDwel EOL Volt Trigger Delays and Phase SynchronizationVoltm Fixed List Count Dwel Volt EOL Step AutoSyncsour Phase Example DisplaySyncphas Initimmed Using Slew Rates to Generate Waveforms Programming Slew RatesStep Slewm Voltm FixedSlew SlewtCurrlev Syncsour Phase SyncphasMeasuring Peak Inrush Current CurrpeakSaving and Recalling Operating States Setting the Gpib Address and RS-232 ParametersTo configure the RS-232 interface, proceed as follows Action Display To set the Gpib address, proceed as followsSpecifications Table A-1. Performance Specifications1Table A-1. Performance Specifications Supplemental Characteristics Table A-2. Supplemental CharacteristicsTrig Out HC TTL output Trig In 10k pullup Operation Below 45 Hz Table A-3. Operation Below 45 HzPage Verification and Calibration Equipment RequiredTest Setup Table B-1. Equipment RequiredRecommended Model AC Voltage Programming and Measurement Accuracy Turn-On Checkout ProcedurePerforming the Verification Tests Action Normal ResultVolt 0, Outpcoup DC, Offset DC Voltage Programming and Measurement AccuracyRMS Current Accuracy Test Volt 50, Currrange LOW, CurrlevDC Programming and Measurement Accuracy Table B-2. Agilent 6811B Verification Test RecordTable B-3. Agilent 6812B Verification Test Record RMS Current Measurement AccuracyTable B-3 Table B-4. Agilent 6813B Verification Test RecordPerforming the Calibration Procedure Front Panel Calibration MenuEnable Calibration Mode Front Panel CalibrationCalibrating and Entering Voltage Offset Values Calibrating and Entering DC Voltage Gain Values Calibrating and Entering AC rms Voltage Gain ValuesCallev P4 Calibrating the OVP Trip Point Calibrating and Entering rms Current ValuesSaving the Calibration Constants Calibrating and Entering rms Current Measurement ValuesCalibrating the Output Impedance OFFTable B-5. Gpib Calibration Error Messages Changing the Calibration PasswordCalibration Error Messages Calibration Over the GpibFigure B-2. Calibration Program Listing Sheet 1 Figure B-2. Calibration Program Listing Sheet 2 Figure B-2. Calibration Program Listing Sheet 3 Page Error Number List Table C-1. Error NumbersError Messages Error Messages C Page Open the Unit Check the Jumper Wire Model Agilent 6811B/6812B onlyCheck the Power Transformer Connector all Models Check the Line Jumpers all ModelsClose the Unit Figure D-1. Line Voltage Conversion ComponentsIndex Index Index Europe Asia Pacific United States Latin AmericaCanada Australia/New Zealand JapanManual Updates
Related manuals
Manual 187 pages 38.86 Kb Manual 20 pages 27.01 Kb Manual 3 pages 27.44 Kb

6812B, 6811B, 6813B specifications

Agilent Technologies, a leader in electronic test and measurement equipment, offers a range of powerful signal sources including the 6843A, 6834B, 6814B, 6813B, and 6811B models. These instruments are designed to support various applications in research, development, and manufacturing, providing precise signal generation capabilities.

The Agilent 6843A is a versatile signal generator known for its exceptional frequency range and modulation capabilities. It supports an extensive bandwidth, making it ideal for applications that require high-frequency signal generation. With its superior phase noise performance, the 6843A is an excellent choice for radar, wireless communications, and electronic warfare applications. The instrument features an intuitive user interface, allowing engineers to set parameters quickly and efficiently.

Next, the Agilent 6834B offers exceptional performance characteristics, including high output power and low distortion. This signal generator is particularly noted for its ability to produce complex modulation formats, making it suitable for testing advanced wireless communication systems. With a reliable and stable output, the 6834B ensures accurate and repeatable measurements, which is vital for thorough testing processes.

The 6814B model stands out for its dual-channel capabilities, allowing users to generate simultaneous signals for testing multiple components or systems. This feature significantly enhances testing efficiency and flexibility for engineers. With built-in arbitrary waveform functionality, users can create custom waveforms, making the 6814B suitable for a wide range of applications including device characterization and signal processing research.

For those seeking a more compact solution, the Agilent 6813B provides essential signal generation features without compromising on performance. It is designed for a variety of applications across telecommunications and consumer electronics, featuring a straightforward interface and robust performance metrics.

Lastly, the 6811B is an entry-level yet capable model that supports a broad spectrum of testing needs. Perfect for educational and laboratory environments, it provides essential functionalities required for effective signal generation and analysis.

Overall, Agilent Technologies' 6843A, 6834B, 6814B, 6813B, and 6811B signal generators offer an array of features and technologies that cater to various application needs. Their precision, reliability, and user-oriented designs position them as invaluable assets in any testing environment, ensuring engineers can carry out their work with confidence and accuracy.