Agilent Technologies Essential Operation Instructions for Agilent Yard Vacuum Models

Page 56

4 - Front Panel Operation

 

Example

Display

ϕ

This example uses the immediate slew rate. First, access the Voltage menu

VOLT:M

 

FIXED

 

and press until you access the mode command. On the Entry keypad, press

 

 

 

 

° to obtain FIXED. Press Enter.

 

 

 

 

Access the voltage menu and press until you access the slew command. On

SLEW

50

 

the Entry keypad, press 5 0 Enter to program a slew rate of 50 volts/second.

 

 

 

 

Whenever a new immediate voltage value is entered, the output will slew to

 

 

 

 

the new level at 50 volts/second.

 

 

 

Step mode uses the triggered slew rate. First, access the Voltage menu,

VOLT

 

120

 

program the immediate and triggered voltage levels, and set the slew mode to

VOLT:T

150

 

STEP.

SLEW:M

STEP

 

Access the Voltage menu and press to access the immediate slew

SLEW:

9.9+E37

 

command. On the Entry keypad, enter a value that equals infinity.

 

 

 

 

Access the Voltage menu and press until you access the triggered slew

SLEW:T

50

 

command. On the Entry keypad, enter a value such as 5 0 Enter, which sets

 

 

 

 

the triggered slew rate to 50 volts/second.

 

 

 

 

Then press Trigger Control and Enter, followed by Shift Trigger.

INIT:IMMED

 

After the trigger has been sent, in step mode, the triggered value becomes the

 

 

 

 

new immediate value.

 

 

 

Pulse mode uses the triggered slew rate at the leading edge of the pulse, and

VOLT

 

120

 

the immediate slew rate at the trailing edge of the pulse. First, access the

VOLT:T

150

 

Voltage menu, program the immediate and triggered voltage levels, and set

SLEW:M

 

PULSE

 

the slew mode to PULSE.

 

 

 

 

Access the Pulse menu and program the pulse count, duty cycle, and pulse

COUNT 2

 

period.

DCYCLE

33

 

 

PER

.0166

 

Access the Voltage menu and press to access the immediate slew

SLEW:

9.9+E37

 

command. On the Entry keypad, enter a value that equals infinity.

 

 

 

 

Access the Voltage menu and press until you access the triggered slew

SLEW:T

50

 

command. On the Entry keypad, enter a value such as 5 0 Enter, which sets

 

 

 

 

the triggered slew rate to 50 volts/second.

 

 

 

 

Then press Trigger Control and Enter, followed by Shift Trigger.

INIT:IMMED

When the voltage slew mode is set to LIST, the slew rates are set by the

 

 

 

values in the voltage slew list. Refer to the List Transient example for more

 

 

 

information on how to program lists. You must program the voltage values and dwell times as explaied in that example. You must also program a slew rate for each point in the list (even if it is 9.9+E37).

NOTE: When specifying a dwell time, you must take the slew time into consideration. If the dwell time at any given list point is less than the slew time at the same point, the voltage will never reach its programmed level before the next list point becomes active.

56

Image 56
Contents AC Power Solutions Agilent Models 6811B, 6812B, and 6813B User’s GuideWarranty Information CertificationSafety Summary GeneralSafety Symbols Declaration EMCAcoustic Noise Information Printing HistoryTable of Contents Performing the Calibration Procedure Entry Keys Examples of Front Panel ProgrammingIntroduction Performing the Verification TestsError Number List Line Voltage ConversionPage Location Document OrientationTopic Safety Considerations Options and PartsEarlier AC Source Models OptionsAgilent Part Number DescriptionOperator Replaceable Parts Model Description Front Panel/Remote OperationCapabilities Steady-state Output Characteristic Steady-state Output Characteristic in real-time modePeak Current/Dynamic Power Capability Peak Current LimitPeak Inrush Example Typical Peak Current Output CapacitiesIpeak setting Capacitance in μF 127 254RMS Current Limit Circuit Voltage RegulationOutput Impedance Real Time RegulationOutput Coupling Page Cleaning InspectionDamage Packaging Material Items SuppliedLocation Bench OperationRack Mounting Not block the fan exhaust at the rear of the unitInput Source and Line Fuse Installing the Power CordInput Connections Connecting the Power Cord Output ConnectionsWire Considerations Remote Sense ConnectionsVoltage Drops Current RatingsRemote Sense Connections Remote Sensing and OVP Considerations Trigger ConnectionsDigital Connections Gpib Connector Controller ConnectionsInterface Commands RS-232 ConnectorRS-232 Interface RS-232 Data FormatHardware Handshake Null Modem Interface Lines Response Data TerminatorIntroduction Preliminary CheckoutUsing the Keypad Checkout ProcedurePress 1, 2, 0, Enter Press Output On/OffPress Protect Press 1, 6, 0, EnterError Messages Selftest ErrorsPower-On Selftest Errors Case of TroubleRuntime Error Messages Line FuseIndicates May appear at runtimePage Front Panel Description Front Panel, Overall ViewAC+DC Dis lay Command Function System KeysScrolling Keys Function KeysImmediate Action Keys Display Command Function Meter Display KeysDisplay Measurement DisplayOutput Control Keys ListRST CLS Protection and Status Control KeysProtclear Trigger and List Control Keys InitimmedIMM AbortEntry Keys Through Are used for entering numeric values Is the decimalSet the output to 120 V rms as follows Setting the Output Voltage AmplitudeExamples of Front Panel Programming Setting the Output Frequency Setting the DC OffsetTo verify the output, use the meter menu ActionSetting a Protection Feature Overcurrent protection feature as followsBit Number Bit Weight Clearing Protection ConditionsUsing Transient Voltage Modes PulseStep Transient Pulse TransientVoltt Voltm StepList Transient Voltm PulseWidth DcycleTrigger Delays and Phase Synchronization Voltm Fixed List Count DwelDwel EOL Volt Volt EOL Step AutoSyncphas Initimmed Syncsour PhaseExample Display Using Slew Rates to Generate Waveforms Programming Slew RatesVoltm Fixed SlewStep Slewm SlewtSyncsour Phase Syncphas Measuring Peak Inrush CurrentCurrlev CurrpeakSetting the Gpib Address and RS-232 Parameters To configure the RS-232 interface, proceed as followsSaving and Recalling Operating States Action Display To set the Gpib address, proceed as followsSpecifications Table A-1. Performance Specifications1Table A-1. Performance Specifications Supplemental Characteristics Table A-2. Supplemental CharacteristicsTrig Out HC TTL output Trig In 10k pullup Operation Below 45 Hz Table A-3. Operation Below 45 HzPage Verification and Calibration Equipment RequiredRecommended Model Test SetupTable B-1. Equipment Required Turn-On Checkout Procedure Performing the Verification TestsAC Voltage Programming and Measurement Accuracy Action Normal ResultDC Voltage Programming and Measurement Accuracy RMS Current Accuracy TestVolt 0, Outpcoup DC, Offset Volt 50, Currrange LOW, CurrlevTable B-2. Agilent 6811B Verification Test Record Table B-3. Agilent 6812B Verification Test RecordDC Programming and Measurement Accuracy RMS Current Measurement AccuracyTable B-3 Table B-4. Agilent 6813B Verification Test RecordPerforming the Calibration Procedure Front Panel Calibration MenuCalibrating and Entering Voltage Offset Values Enable Calibration ModeFront Panel Calibration Callev P4 Calibrating and Entering DC Voltage Gain ValuesCalibrating and Entering AC rms Voltage Gain Values Calibrating the OVP Trip Point Calibrating and Entering rms Current ValuesCalibrating and Entering rms Current Measurement Values Calibrating the Output ImpedanceSaving the Calibration Constants OFFChanging the Calibration Password Calibration Error MessagesTable B-5. Gpib Calibration Error Messages Calibration Over the GpibFigure B-2. Calibration Program Listing Sheet 1 Figure B-2. Calibration Program Listing Sheet 2 Figure B-2. Calibration Program Listing Sheet 3 Page Error Number List Table C-1. Error NumbersError Messages Error Messages C Page Check the Jumper Wire Model Agilent 6811B/6812B only Check the Power Transformer Connector all ModelsOpen the Unit Check the Line Jumpers all ModelsClose the Unit Figure D-1. Line Voltage Conversion ComponentsIndex Index Index United States Latin America Canada Australia/New ZealandEurope Asia Pacific JapanManual Updates
Related manuals
Manual 187 pages 38.86 Kb Manual 20 pages 27.01 Kb Manual 3 pages 27.44 Kb

6812B, 6811B, 6813B specifications

Agilent Technologies, a leader in electronic test and measurement equipment, offers a range of powerful signal sources including the 6843A, 6834B, 6814B, 6813B, and 6811B models. These instruments are designed to support various applications in research, development, and manufacturing, providing precise signal generation capabilities.

The Agilent 6843A is a versatile signal generator known for its exceptional frequency range and modulation capabilities. It supports an extensive bandwidth, making it ideal for applications that require high-frequency signal generation. With its superior phase noise performance, the 6843A is an excellent choice for radar, wireless communications, and electronic warfare applications. The instrument features an intuitive user interface, allowing engineers to set parameters quickly and efficiently.

Next, the Agilent 6834B offers exceptional performance characteristics, including high output power and low distortion. This signal generator is particularly noted for its ability to produce complex modulation formats, making it suitable for testing advanced wireless communication systems. With a reliable and stable output, the 6834B ensures accurate and repeatable measurements, which is vital for thorough testing processes.

The 6814B model stands out for its dual-channel capabilities, allowing users to generate simultaneous signals for testing multiple components or systems. This feature significantly enhances testing efficiency and flexibility for engineers. With built-in arbitrary waveform functionality, users can create custom waveforms, making the 6814B suitable for a wide range of applications including device characterization and signal processing research.

For those seeking a more compact solution, the Agilent 6813B provides essential signal generation features without compromising on performance. It is designed for a variety of applications across telecommunications and consumer electronics, featuring a straightforward interface and robust performance metrics.

Lastly, the 6811B is an entry-level yet capable model that supports a broad spectrum of testing needs. Perfect for educational and laboratory environments, it provides essential functionalities required for effective signal generation and analysis.

Overall, Agilent Technologies' 6843A, 6834B, 6814B, 6813B, and 6811B signal generators offer an array of features and technologies that cater to various application needs. Their precision, reliability, and user-oriented designs position them as invaluable assets in any testing environment, ensuring engineers can carry out their work with confidence and accuracy.