Calibration Table
CAUTION
Dangerous voltages are present during the calibration procedure. Calibration should only be performed by qualified service technicians using a
Step | Function | Range | Input Signal | Adjustment | Tolerance | |
(limit) | (counts) | |||||
|
|
|
| |||
1 |
|
| Short | — | ±2 | |
2 | 500 mV | 500 mV | 480.0 mV | 1 (±2) | ±26 | |
3 |
|
| — | ±26 | ||
4 |
| 50 V | 48.000 V | 2 (±2) | ±26 | |
5 |
| — | ±26 | |||
|
| |||||
6 | V | 5 V | 4.800 V | 3 (±2) | ±26 | |
7 |
| 500 V | 480.00 V | RV4 (±2) | ±26 | |
8 |
| 1000 V | 1000 V | RV5 (±1) | ±7 | |
9* | V | 5 V | 4.8000 V @ 200 Hz | 6 (±10) | ±270 | |
10* | 0.2500 V @ 200 Hz | 7 (±5) | ±42 | |||
|
| |||||
11 |
| 500 mV | 480.0 mV @ 200 Hz | 8 (±10) | ±366 | |
12 |
|
| 480.00 V @ 10 kHz | C1 (±20) | ±270 | |
13 |
| 500 V | 480.00 V @ 100 kHz | — | ±1740 | |
14 |
|
| 480.00 V @ 200 Hz | — | ±270 | |
15 |
| 5 V | 4 .8000 V @ 10 kHz | C2 (±20) | ±270 | |
16 |
| 4 .8000 V @ 100 kHz | — | ±1740 | ||
|
|
*Repeat steps 9 and 10.