HP Video Game 00974-90002 manual Maximum Overvoltage Limitations AC and DC Voltage Functions

Page 5

Safety Summary

WARNING

POSSIBLE ELECTRICAL SHOCK. Calibration and performance tests are to be performed by qualified personnel only. Do not attempt calibration or test procedures unless qualified to do so.

CAUTION

To avoid damage to the multimeter for inputs above 250 Vdc or Vac, disconnect the test leads before changing functions. Do not exceed the maximum input limits.

Maximum Overvoltage Limitations (AC and DC Voltage Functions)

1000V

MAX indicates the maximum voltage between input terminals and earth is ± 1000 V (dc or ac rms).

Do not use the multimeter on any ACV circuit where the maximum impulse overvoltage may be more than 4000Vpk or any DCV circuit where the maximum impulse overvoltage may be more than 2500Vpk between the COM and VOLT terminals. Excessive impulse overvoltage can damage the multimeter voltage functions. Do not measure branch circuits (CAT II) over 600V to earth or service panel circuits (CAT III) over 300V to earth.

1-5

Image 5
Contents HP 974A Multimeter User’s Guide Page HP 974A Multimeter Table of ContentsSafety Symbols Safety SummaryMaximum Overvoltage Limitations AC and DC Voltage Functions Safety SummaryFunction Maximum Operating Input Probes and Test LeadsTerminals, Shutter, & Test Leads OperationSwitch Display Position Function SwitchPress Action Main Display Secondary Display Function KeysAction Main Display Secondary Display Minimum/MaximumHold/Auto-Hold Data Hold Auto-Hold Range Percent Function Keys and Function Switch MatrixFunction Relative Min/Max Auto Power Off Power onDisplay AudioRequired Test Equipment Calibration and AdjustmentCalibration Procedure Fuse Replacement MaintenanceBattery Replacement Cleaning TroubleshootingReplaceable Parts/Accessories Problem Possible Cause Suggested ActionRange Resolution Accuracy Input Resistance SpecificationsGeneral DC VoltageDC Current, AC Current 40 Hz to 1 kHz, 5% to 100% of range AC + DC Voltage rms responding, computed from acV, dcVContinuity ResistanceDiode Frequency AC CoupledTemperature 5 kΩ @ 25C Thermistor probe Adjustments Limit Counts Step Function Range Input Signal AdjustmentCalibration Table Step Function Range Input Signal Call out Description HP Part Number Replaceable Parts/AccessoriesDisassembly Manufacturer’s Name Conforms to the following Product SpecificationsManufacturer’s Address Declares, that the productsWhat is Not Covered Warranty/ServiceWhat is Covered