Baud rate from 300 to 57600 No parity
8 data bits
1 stop bit.
Your output device must be configured in the same way in order to communicate with ColorFlex. Note that Universal Software’s maximum baud rate is 19200.
Note: Several baud rates are available with ColorFlex. Be sure to change the INSTRUMENT SETUP of ColorFlex to agree with the baud rate selected for your output device.
Handshake
Output from the ColorFlex serial port is controlled by XON/XOFF handshaking. Data output is stopped when the output device sends an XOFF character to ColorFlex. Data flow begins again when ColorFlex receives an XON character.
Conditions of Illumination and Viewing
Source | |
Source UV Content | Nominal match to D65 |
Integrating Sphere | 63.5 mm diameter, |
Port Diameters/View Diameters | |
45/0 Model | 31.8 mm (1.25”)/25.4 mm (1.0”) |
Diffuse/8° | 14.3 mm (0.56”)/8.0 mm (0.31”) |
Power Required
External 120/220 V locking AC adapter
Power Cord Wire Color Code:
| Color | Definition |
220V Cord | Brown | LINE |
| Blue | NEUTRAL |
| Green/Yellow SAFETY | |
110V Cord | Black | LINE |
| White | NEUTRAL |
| Green | SAFETY |
Instrument-Computer Ground Potential Check:
Perform this check if the power cord wiring is being changed, such as for replacement of the cord or if changing the plug.