The instrument serial port ground connection is referenced to the instrument's internal frame and safety ground. Before connecting the communication cable to the instrument serial port, apply power to the instrument and the host computer. Check the ground potential (voltage) between the serial port ground pins on the computer and the instrument. The ground connection is pin 5 on DB-9 connectors and pin 7 on DB-25 connectors. Voltages in excess of 5VAC at 110V input power, or in excess of 10VAC at 220V input power may indicate a difference in ground wiring and can damage the instrument and/or the computer. Check the wiring and take other steps as needed to reduce this difference before connecting the communications cable. You may also use a data optoisolator.

Instrument Performance

Spectral Data

Range (nm)

400 - 700

Photometric Range

0 - 150%

Spectral Resolution

10 nm

 

Spectral Bandpass

< 12 nm

Measurement Speed (at 23° C)

≤ 1 second

Warm-up time

 

Command-to-data output

≤ 1 second

Read-to-read

 

≤ 3 seconds

Warm-up Period

30 minutes prior to testing

Sample Storage Capacity

500 spectral samples

 

999 tristimulus samples

 

99 standards

 

99 setups

You may test the repeatability of your instrument as follows:

1.Turn the ColorFlex on and allow it to warm it up for 30 minutes. Meanwhile, clean the black glass (45°/0° instruments) and white tile as described in Chapter 24 and allow the tiles to return to room temperature.

2.Standardize the instrument.

3.Configure your color data display to show DE* using D65/10°.

4.Center the instrument’s white standard tile at the measurement port. Do not move or remove the tile for the remainder of the test.

5.Read the white tile as a standard.

6.Configure the timed read feature to automatically make and average 20 more measurements with a read interval of 7 seconds and then initiate reading of the tile as a sample.

13-20

ColorFlex

07/02

Image 20
Contents ColorFlex Sensor Read ButtonDiffuse model Accessories Options and Sample Devices Sample Clamp/Port-Forward Stand Port Inserts 45/0 instruments onlyInstalling the Bar Code Reader Bar Code Reader KitReading and Saving Bar Code IDs Inch Glass Sample Cup Sample Cup Opaque CoverSample Cup Port Plate 45/0 instruments only Ring and Disk Set 45/0 instruments onlyOrange Juice Tube Holder 45/0 instruments only Citrus Option 45/0 instruments onlyOJ4 Orange Juice Standard Set of 6 Orange Juice StandardsTomato Option 45/0 instruments only Tomato Red TileInch UV Port Insert 45/0 instruments only Skein Holder 45/0 instruments onlyInch Glass Port Insert 45/0 instruments only Foot Switch External Serial Printer Carrying CaseInstallation Standardization 07/02 ColorFlex 13-15 Maintenance Cleaning the ColorFlexReplacing the Lamp Operating ConditionsArchiving Setups Error MessagesResetting the ColorFlex Left Arrow key + Right Arrow key + Read keySpecifications Physical CharacteristicsPower Required Conditions of Illumination and ViewingInstrument Performance Regulatory Notice Supplementary Information Conforms to the following Product SpecificationsAccording to ISO/IEC Guide 22 and EN Color Measurement Notes 13-24 ColorFlex 07/02