7.When all readings are complete, examine the average DE* value. The instrument passes the repeatability test as long as the average DE* value is less than or equal to 0.04. Contact HunterLab Customer Support as described on page 25-6 if the instrument fails the repeatability test.

Note: Performance specifications are subject to change without notice.

Regulatory Notice

A copy of the Declaration of Conformity according to ISO/IEC Guide 22 and EN 45014 follows on the next page.

07/02

ColorFlex

13-21

Image 21
Contents Read Button ColorFlex SensorDiffuse model Accessories Options and Sample Devices Port Inserts 45/0 instruments only Sample Clamp/Port-Forward StandBar Code Reader Kit Installing the Bar Code ReaderReading and Saving Bar Code IDs Sample Cup Opaque Cover Inch Glass Sample CupSample Cup Port Plate 45/0 instruments only Ring and Disk Set 45/0 instruments onlyCitrus Option 45/0 instruments only Orange Juice Tube Holder 45/0 instruments onlySet of 6 Orange Juice Standards OJ4 Orange Juice StandardTomato Option 45/0 instruments only Tomato Red TileSkein Holder 45/0 instruments only Inch Glass Port Insert 45/0 instruments onlyInch UV Port Insert 45/0 instruments only External Serial Printer Carrying Case Foot SwitchInstallation Standardization 07/02 ColorFlex 13-15 Cleaning the ColorFlex MaintenanceReplacing the Lamp Operating ConditionsError Messages Archiving SetupsResetting the ColorFlex Left Arrow key + Right Arrow key + Read keyPhysical Characteristics SpecificationsConditions of Illumination and Viewing Power RequiredInstrument Performance Regulatory Notice Conforms to the following Product Specifications According to ISO/IEC Guide 22 and ENSupplementary Information Color Measurement Notes 13-24 ColorFlex 07/02