HP 3325B, 3325A manual High Voltage Output Option 002 only Performance Check

Page 20

TB 9-6625-2049-35

(4)Enter -4.5 VOLT using DATA keys. Multimeter will indicate between -4.35 and

-4.65 V dc.

(5)Press ENTRY FREQ key and enter 999.9 kHz using DATA keys. Multimeter will indicate between -4.44 and -4.56 V dc.

(6)Press ENTRY DC OFFSET key and enter 4.5 VOLT using DATA keys. Multimeter will indicate between 4.44 and 4.56 V dc.

(7)Press FUNCTION square wave key. Multimeter will indicate between 4.44 and

4.56V dc.

(8)Enter -4.5 VOLT using DATA keys. Multimeter will indicate between -4.44 and

-4.56 V dc.

(9)Press ENTRY FREQ key and enter 9.9999 MHz using DATA keys. Multimeter will indicate between -4.35 and -4.65 V dc.

(10)Press FUNCTION triangle key and enter 9.9 kHz using DATA keys.

(11)Multimeter will indicate between -4.44 and -4.56 V dc.

(12)Press FUNCTION positive ramp key. Multimeter will indicate between -4.38 and -4.62 V dc.

b.Adjustments. No adjustments can be made.

19.High Voltage Output Option 002 only

a.Performance Check

(1) Connect SIGNAL to multimeter INPUT HI and LO.

(2) Press keys and enter values using DATA keys as listed in (a) through (i) below:

(a)FUNCTION sine wave.

(b)ENTRY FREQ.

(c)2 kHz.

(d)SIGNAL on.

(e)ENTRY DC OFFSET.

(f)0 VOLT.

(g)ENTRY AMPTD.

(h)14.14 V RMS.

(i)AMPTD CAL.

(3)Multimeter will indicate between 13.86 and 14.42 V ac.

(4)Press keys and enter values using DATA keys as listed in (a) through (e) below:

20

Image 20
Contents Models 3325A and 3325B Section Identification and Description Forms, Records, and ReportsProgrammed amplitude at 1 kHz Section Equipment Requirements Spectrum Analyzer Preliminary Instructions Equipment SetupSection Calibration Process Entry Amptd 20 dBm Entry DC Offset Volt Entry Freq 20 MHz Harmonic Distortion a. Performance CheckIN/OUT Signal on Entry Freq 25 kHz Entry Amptd Volt 50 Hz Entry Amptd VoltEntry Freq MHz Entry Amptd Volt Entry DC Offset Square Wave Characteristics Performance CheckAmplitude Modulation Distortion a. Performance Check Entry Freq 10 kHz Entry Amptd Volt Entry Freq MHz Entry Amptd RMSRamp Retrace Performance Check Entry Freq 60 MHz Frequency Accuracy Performance CheckEntry Freq 20 MHz Entry Amptd Volt Adjustments SIG InputEntry Freq 100 kHz Entry Amptd 13 dBm Entry PhaseBlue then Asgn Zero Phase Entry Amptd RMS Entry Freq 100 Hz Amptd CAL Sine wave Amplitude and Flatness Performance CheckEntry DC Offset Volt Entry DC Offset Entry Freq KHz Entry Amptd RMS Amptd CAL Press Entry Amptd keySquare Wave Amplitude and Flatness Performance Check KHz Entry Amptd RMS Amptd CAL Triangle and Ramp Amplitude Performance Check100 Hz Amptd CAL Dc Offset Performance CheckDc Offset with Ac Performance Check High Voltage Output Option 002 only Performance Check Entry Freq KHz Entry Amptd RMS Amptd CAL Power SupplyEntry Freq 100 kHz Entry Amptd RMS Amptd CAL Final Procedure Power supply assembly A2 test instrument top viewPIN