HP 3325B, 3325A manual Hz Amptd CAL, Dc Offset Performance Check

Page 18

TB 9-6625-2049-35

(6)Enter 10 kHz using DATA keys. Multimeter will indicate between 2.742 and 3.032 V ac.

(7)Press keys and enter values using DATA keys as listed in (a) through (c) below:

(a)FUNCTION positive ramp.

(b)100 Hz.

(c)AMPTD CAL.

(8)Multimeter will indicate between 2.843 and 2.931 V ac.

(9)Enter 499 Hz using DATA keys. Multimeter will indicate between 2.843 and 2.931 V ac.

(10)Enter 501 Hz using DATA keys. Multimeter will indicate between 2.598 and 3.176 V ac.

(11)Enter 10 kHz using DATA keys. Multimeter will indicate between 2.598 and

3.176V ac.

b.Adjustments. No adjustments can be made.

17.Dc Offset

a.Performance Check

(1) Connect TI SIGNAL to multimeter INPUT HI and LO using 50Ω feedthrough

termination.

(2)Press SIGNAL key off (option 002 not lit).

(3)Press presently active FUNCTION key to remove ac output and activate ENTRY DC OFFSET key (lit).

(4)Enter 5 VOLT using DATA keys and press AMPTD CAL key.

(5)Multimeter will indicate between 4.98 and 5.02 V dc.

(6)Enter -5 VOLT using DATA keys. Multimeter will indicate between -4.98 and -5.02 V dc.

(7)Repeat technique of (6) above using TI settings and multimeter indications listed in table 5.

Table 5. Dc Offset

Test instrument

Multimeter indications

DC OFFSET

 

(V dc)

settings

 

Min

 

Max

1.499

VOLT

1.493

 

1.505

-1.499

VOLT

-1.505

 

-1.493

499.9

mV

0.4979

 

0.5019

-499.9

mV

-0.5019

 

-0.4979

149.9

mV

0.1493

 

0.1505

-149.9

mV

-0.1505

 

-0.1493

49.99

mV

0.04979

 

0.05019

-49.99

mV

-0.05019

 

-0.04979

18

Image 18
Contents Models 3325A and 3325B Section Identification and Description Forms, Records, and ReportsProgrammed amplitude at 1 kHz Section Equipment Requirements Spectrum Analyzer Equipment Setup Section Calibration ProcessPreliminary Instructions Harmonic Distortion a. Performance Check IN/OUTEntry Amptd 20 dBm Entry DC Offset Volt Entry Freq 20 MHz Signal on Entry Freq 25 kHz Entry Amptd Volt 50 Hz Entry Amptd VoltSquare Wave Characteristics Performance Check Amplitude Modulation Distortion a. Performance CheckEntry Freq MHz Entry Amptd Volt Entry DC Offset Entry Freq MHz Entry Amptd RMS Ramp Retrace Performance CheckEntry Freq 10 kHz Entry Amptd Volt Frequency Accuracy Performance Check Entry Freq 20 MHz Entry Amptd VoltEntry Freq 60 MHz Adjustments SIG InputEntry Phase Blue then Asgn Zero PhaseEntry Freq 100 kHz Entry Amptd 13 dBm Sine wave Amplitude and Flatness Performance Check Entry DC Offset VoltEntry Amptd RMS Entry Freq 100 Hz Amptd CAL Entry DC Offset Entry Freq KHz Entry Amptd RMS Amptd CAL Press Entry Amptd keySquare Wave Amplitude and Flatness Performance Check KHz Entry Amptd RMS Amptd CAL Triangle and Ramp Amplitude Performance Check100 Hz Amptd CAL Dc Offset Performance CheckDc Offset with Ac Performance Check High Voltage Output Option 002 only Performance Check Power Supply Entry Freq 100 kHz Entry Amptd RMS Amptd CALEntry Freq KHz Entry Amptd RMS Amptd CAL Final Procedure Power supply assembly A2 test instrument top viewPIN