HP 3325A Entry Freq 100 kHz Entry Amptd RMS Amptd CAL, Entry Freq KHz Entry Amptd RMS Amptd CAL

Page 21

TB 9-6625-2049-35

(a)ENTRY FREQ.

(b)100 kHz.

(c)ENTRY AMPTD.

(d)10 V RMS.

(e)AMPTD CAL.

(5)Multimeter will indicate between 9 and 11 V ac.

(6)Press keys and enter values using DATA keys as listed in (a) through (f) below:

(a)FUNCTION square wave.

(b)ENTRY FREQ.

(c)2 kHz.

(d)ENTRY AMPTD.

(e)20 V RMS.

(f)AMPTD CAL.

(7)Multimeter will indicate between 19.6 and 20.4 V ac.

(8)Press FUNCTION triangle key and enter 11.5 V RMS using DATA keys.

(9)Press AMPTD CAL key. Multimeter will indicate between 11.27 and 11.73 V ac.

(10)Press FUNCTION positive ramp key and repeat (9) above. b. Adjustments. No adjustments can be made.

20.Power Supply

NOTE

Do not perform power supply check if all other parameters are within tolerance.

a.Performance Check

(1)Connect multimeter INPUT HI to -15 V TP (fig. 5) and LO to ground side of C9 (fig. 5). Adjust R22 (fig. 5) for a multimeter indication of -14.97 to -15.03 V dc (R).

(2)Connect multimeter INPUT HI to +15 V TP (fig. 5). Readjust R22 (fig. 5) if multimeter does not indicate between 14.9 and 15.1 V dc.

(3)Connect multimeter INPUT HI to +5 V TP (fig. 5). Readjust R22 (fig. 5) if multimeter does not indicate between 5.01 and 5.05 V dc.

(4)Repeat (1) through (3) above for best in-tolerance condition on all test points.

21

Image 21
Contents Models 3325A and 3325B Forms, Records, and Reports Section Identification and DescriptionProgrammed amplitude at 1 kHz Section Equipment Requirements Spectrum Analyzer Equipment Setup Section Calibration ProcessPreliminary Instructions Harmonic Distortion a. Performance Check IN/OUTEntry Amptd 20 dBm Entry DC Offset Volt Entry Freq 20 MHz 50 Hz Entry Amptd Volt Signal on Entry Freq 25 kHz Entry Amptd VoltSquare Wave Characteristics Performance Check Amplitude Modulation Distortion a. Performance CheckEntry Freq MHz Entry Amptd Volt Entry DC Offset Entry Freq MHz Entry Amptd RMS Ramp Retrace Performance CheckEntry Freq 10 kHz Entry Amptd Volt Frequency Accuracy Performance Check Entry Freq 20 MHz Entry Amptd VoltEntry Freq 60 MHz SIG Input AdjustmentsEntry Phase Blue then Asgn Zero PhaseEntry Freq 100 kHz Entry Amptd 13 dBm Sine wave Amplitude and Flatness Performance Check Entry DC Offset VoltEntry Amptd RMS Entry Freq 100 Hz Amptd CAL Press Entry Amptd key Entry DC Offset Entry Freq KHz Entry Amptd RMS Amptd CALSquare Wave Amplitude and Flatness Performance Check Triangle and Ramp Amplitude Performance Check KHz Entry Amptd RMS Amptd CALDc Offset Performance Check 100 Hz Amptd CALDc Offset with Ac Performance Check High Voltage Output Option 002 only Performance Check Power Supply Entry Freq 100 kHz Entry Amptd RMS Amptd CALEntry Freq KHz Entry Amptd RMS Amptd CAL Power supply assembly A2 test instrument top view Final ProcedurePIN