MODEL GFC7000EU CO ANALYZER WITH AUTO-REFERENCE

(Addendum to GFC7000E Manual PN 04584)

 

Humidity Range

0-95% RH, Non-Condensing

 

Temp Coefficient

< 0.05 % of reading per °C (5 ppb/°C minimum)(6)

 

Voltage Coefficient

< 0.05 % of reading per V

 

Dimensions

7" x 17" x 23.5" (178 mm x 432 mm x 597 mm)

 

(HxWxD)

 

 

Weight

50 lb (22.7 kg)

 

AC Power

100V 50/60 Hz (3.25A), 115 V 60 Hz (3.0A),

 

 

220 – 240 V 50/60 Hz (2.5A)

 

Environmental

Installation Category (Over voltage Category) II

 

Conditions

Pollution Degree 2

 

Analog Outputs

Three (3) Outputs

 

Analog Output

100 mV, 1 V, 5 V, 10 V, 2-20 or 4-20 mA isolated

 

Ranges

current loop.

 

 

All Ranges with 5% Under/Over Range

 

Analog Output

1 part in 4096 of selected full-scale voltage

 

Resolution

 

 

Status Outputs

8 Status outputs from opto-isolators

 

Control Inputs

6 Control Inputs, 2 defined, 4 spare

 

I/O

One (1) RS-232; One (1) RS-485/RS-232/Ethernet

 

 

Baud Rate : 300 - 115200

 

Certifications

CE: EN61010-1:90 + A1:92 + A2:95, EN61326 - Class

 

 

A

(1)As defined by the USEPA

(2)At constant temperature and voltage

(3)Or 0.2 ppm, whichever is greater

(4)Or 0.1 ppm, whichever is greater

(5)Above 10 ppm range, otherwise 0.2 ppm for lower ranges

3.1.2. Pneumatic Connections

This section is similar to the one in the manual, with the following exception. The extra port labeled “From Purge” is the exhaust port for the for the purge gas. The gas needs to exhausted with a gas line not more than 10 meters long.

In addition, the GFC7000EU uses a different sample pressure sensor than the GFC7000E. The GFc7000EU pressure sensor is rated for oxygen service. The part number is listed in the appendix – spare parts.

6.13.1.3Relay alarm outputs

3

05134 Rev -

Image 3
Contents Teledyne Analytical Instruments Model GFC7000EU CO2 AnalyzerOverview Model GFC7000EU SpecificationsPneumatic Connections MSG Interpreting Warning MessagesCLR Table Warning Messages Fault Condition Possible CausesWarn Fault Condition Possible Causes MessageAppendix B GFC7000EU Spare Parts List #N05ATI Table B-2 GFC7000E Expendables List

GFC7000EU specifications

The Teledyne GFC7000EU is a state-of-the-art gas flow controller renowned for its precision and reliability in various applications. Designed primarily for the semiconductor industry, this advanced instrument excels in managing and controlling gas flows with unparalleled accuracy. Its compact design allows for seamless integration into existing systems, making it a preferred choice for manufacturers aiming to enhance their production processes.

One of the standout features of the GFC7000EU is its ability to provide high-speed control of gas flow. The controller utilizes advanced algorithms that enable rapid response times, ensuring that gas delivery is not only accurate but also stable under dynamic conditions. This is critical in semiconductor manufacturing, where even slight fluctuations in gas flow can lead to defects in the final product.

The GFC7000EU boasts a wide range of flow capabilities, making it versatile for various gases including toxic, corrosive, and reactive types. Its compatibility with multiple gases allows users to optimize their process without the need for extensive modifications to their setups. Additionally, the controller operates efficiently across a broad temperature range, providing reliable performance in varied environmental conditions.

Another significant aspect of the GFC7000EU is its integration with modern communication technologies. It supports various protocols, including RS-232, RS-485, and Ethernet, allowing for easy connection to automation systems and enhancing overall process monitoring and control. This connectivity is vital for real-time data analysis and remote management, improving operational efficiency.

The GFC7000EU is equipped with advanced diagnostics and self-calibration features that minimize downtime. The device includes built-in troubleshooting capabilities and alerts users to any potential issues, ensuring that maintenance can be performed before problems arise. This proactive approach to equipment management enhances productivity and reduces operational costs.

Constructed with high-quality materials, the GFC7000EU is designed for longevity and durability. Its robust housing protects internal components from environmental factors, making it suitable for use in demanding settings. Overall, the Teledyne GFC7000EU stands out as a premier solution for gas flow management, combining cutting-edge technology with user-friendly features to support the evolving needs of the semiconductor and other industries.