Sierra Monitor Corporation 5000 instruction manual SMC sierra monitor corporation

Page 13

SMC sierra monitor corporation

Sentry Instruction Manual - Version 6

7.Apply span gas. Adjust the gain potentiometer until TP10 = correct value as described above. To increase voltage at TP10 turn the gain potentiometer counter clockwise.

8.Remove system power and re-install the sensor and transmitter in the enclosure. Restore power and calibrate.

SERVICE Page: 80

Image 13
Contents N T R Y Cover Plate Combustible Gas Module ServiceCombustible Gas Sensor Wiring Pin Out PIN Function Wire ColorCombustible GAS Scaling Factors Cover Plate Oxygen Module Oxygen sensor should be replaced when it can noSMC sierra monitor corporation Cover Plate Carbon Monoxide Module SMC sierra monitor corporation Cover Plate Hydrogen Sulfide Module Hydrogen Sulfide Transmitter Component Locator Messages Cover Plate Toxic Gas ModuleToxic Gas Transmitter Component Locator SMC sierra monitor corporation Sensor Recharge Cover Plate Ammonia ModuleSMC sierra monitor corporation