SMC sierra monitor corporation | Sentry Instruction Manual - Version 6 |
3.Using a Sierra Monitor Model
4.Remove the calibration gas and wait for the voltage to drop. If the voltage is below 0.30 VDC the procedure is complete. If the voltage is above 0.30 VDC repeat the zero and span adjustments.
5.Calibrate the sensor using the instruction manual procedure.
•A LOW SENSITIVITY message during calibration indicates one of the two conditions for a combustible sensor:
•The zero gas voltage must be less than
.3 volts.
•The new span change (span voltage less zero voltage) must exceed 30% of the calculated span change.
For Example:
Zero gas voltage= .20 VDC Span voltage = .43 VDC
Span change
As the calculated span value (from step 5 above) is 1.03, the calculated span change is:
By calculation: 30% of .87 = .26
The new span change (.23) is less than 30% of the calculated span change so a
LOW SENSITIVITY message will be
displayed.
The LOW SENSITIVITY calculation is made to insure that the calibration does not adjust the electronics so that the "gain" is large enough to cause false alarms due to minor drift or due to minimal electronic noise.
7.5.4 SENSOR REPLACEMENT
If any of the above adjustments cannot be completed successfully the sensor and/or electronics will require replacement. To check the sensor unplug it from the electronics and remove from the housing so that continuity and resistance tests can be made. Table
PIN | FUNCTION | WIRE COLOR |
|
|
|
1 | DETECTOR | WHITE |
|
|
|
2 | COMPENSATOR | BROWN |
|
|
|
3 | CALIB. SW. | RED |
|
|
|
4 | DETECTOR | BLACK |
|
|
|
5 | DRAIN | BARE |
|
|
|
6 | CALIB. SW. | GREEN |
|
|
|
Table
Combustible Gas Sensor
Wiring Pin Out
1.Detector element resistance should be between 1.5 to 2 ohms.
2.Compensator element resistance should be between 1.5 to 2 ohms.
3.Calibration switch should be open in normal operation and should close when the calibration magnet is applied.
If the sensor fails any of these tests it should be replaced. If the sensor passes these tests the electronics are suspect and BOTH the sensor and electronics should be replaced. (The electronics and sensor which are removed should be returned to the factory for evaluation).
When a new sensor is installed the following actions must be taken:
•Make the bridge voltage adjustment immediately to avoid
•Use diagnostic code 0020 to establish nominal value calibration.
•Allow 24 hours for full stabilization of the sensor, recheck the zero and span adjustments and calibrate the sensor module.