Unscrew the old sensor from the conduit hub, screw in the new sensor and connect the wiring harness to the transmitter electronics. Replace the transmitter into the enclosure, make signal adjustments as described above, and replace the enclosure cover.

After the sensor is installed:

Update "new sensor" status in the change module mode.

Allow one hour of stabilization of the new sensor.

Make the electrical sensitivity adjustment as described above.

Calibrate the sensor module.

SERVICE Page: 73

Image 6
Contents N T R Y Service Cover Plate Combustible Gas ModulePIN Function Wire Color Combustible Gas Sensor Wiring Pin OutCombustible GAS Scaling Factors Oxygen sensor should be replaced when it can no Cover Plate Oxygen ModuleSMC sierra monitor corporation Cover Plate Carbon Monoxide Module SMC sierra monitor corporation Cover Plate Hydrogen Sulfide Module Hydrogen Sulfide Transmitter Component Locator Cover Plate Toxic Gas Module MessagesToxic Gas Transmitter Component Locator SMC sierra monitor corporation Cover Plate Ammonia Module Sensor RechargeSMC sierra monitor corporation