Emerson PN 51-1055pHC/rev.K instruction manual Section Description and Specifications

Page 10

MODEL SOLU COMP II

SECTION 1.0

 

DESCRIPTION AND SPECIFICATIONS

SECTION 1.0.

DESCRIPTION AND SPECIFICATIONS

1.1FEATURES AND APPLICATIONS

1.2SPECIFICATIONS

1.3ORDERING INFORMATION AND ACCESSORIES

1.1 FEATURES AND APPLICATIONS

The Solu Comp II analyzers offer the choice of sin- gle or dual sensor input with measurement choices of pH/ORP, resistivity/conductivity/TDS, % concen- tration, ratio conductivity, total and free chlorine, dis- solved oxygen, dissolved ozone, flow and tempera- ture. Dual measurement analyzers offer a wide choice of measurement combinations thus reducing the cost per loop and needed panel space.

FIELD COMMISSION OPTION: The Solu Comp II can be ordered with the ability to commission meas- urements in the field. This added flexibility can greatly reduce the number of spare instruments required for field servicing.

QUICK START PROGRAMMING: Exclusive Quick Start screens appear the first time the Solu Comp II is powered up. Screen prompts direct the user to register the number of sensors, the measurement unit(s) and the language to display. Some measure- ment specific prompts are also displayed. The measurement loop is ready for use in a matter of minutes.

MENUS: Menu screens for calibrating and register- ing choices are simple and intuitive. Plain language prompts guide the user through the procedures.

There are no service codes to enter before gaining access to menus.

DUAL SENSOR INPUT AND OUTPUT: The Solu Comp II accepts single or dual sensor input. The two 4-20 mA outputs can be independently pro- grammed to correspond to any selected measure- ment or temperature. Output damping and linear or log output may also be field selected.

ALARMS: The Solu Comp II has three fully pro- grammable alarm relays that can be assigned to any selected measurement or temperature. Alarms can be configured as high, low, or USP1. The third relay has the additional choice of fault alarm opera- tion. When selected, a fault alarm will activate the relay when a sensor or analyzer fault occurs.

ENCLOSURE: The panel mount version fits stan- dard ½ DIN panel cutouts, and its shallow depth is ideally suited for easy mounting in Hoffman-type enclosures. A panel mount gasket is included to maintain the weather rating of the panel. Surface/pipe mount enclosure includes self-tapping screws for surface mounting. A pipe mounting accessory kit is available for mounting to a 2-inch pipe.

DISPLAY: The two-line, 16-character, back-lit dis- play can be customized to meet user requirements. All operations and descriptive messages can be field selected for English, French, German, Italian, Spanish, or Portuguese. Informative screens, which permit data not shown in the regular display, may be seen at the push of a button.

TEMPERATURE: Most measurements (except ORP and flow) require temperature compensation. The Solu Comp II will automatically recognize either a Pt100 or Pt1000 RTD, normally built into the sen- sor. When this RTD is present, the Solu Comp II can be set up to display the temperature in °C or °F as well as set any one or more of the alarms and/or outputs to respond to this sensor input. If two meas- urements with temperature are present either can be chosen for each alarm and output selected.

1USP alarm applies to conductivity/resistivity only.

1

Image 10
Contents Model Solu Comp Essential Instructions Quick Start Guide ORP Quick Reference Guide Model Solu Comp Table of Contents CONT’D List of TablesList of Figures IiiAbout This Document Section Description and Specifications Hazardous Location DimensionsTerminal Connections Rating 26-14 AWG wire size Ordinary Location -68 onlyRecommended Sensors for Conductivity Performance SpecificationsAnalyzer Conductivity Input Range, µS/cm Cell constant, /cmToroidal Conductivity Codes -21 and/or Measurement Range see table belowWhen used with conductivity -20-32 or Performance Specifications Analyzer ORP InputTemperature Correction none required Recommended Sensors for ORPFrequency Range 0.5 4000 Hz Recommended SensorsRepeatability ±1% Input filter 1- 255 samples Loop Specifications with a Model 499A OZ SensorRecommended Sensor Measurement Range 0-10 ppm mg/LOrdering Information Model 1055 Solu Comp II AnalyzerCombination Guide Part Number Description Weight Unpacking and Inspection Section InstallationInstallation General InformationPanel Mount Installation Panel MountingPipe Mount Installation Pipe MountingSurface Mount Installation Surface MountingPOWER, ALARM, OUTPUT, and Sensor Connections Section WiringPreparing Conduit Openings Model Solu Comp Section Wiring Model Solu Comp Section Wiring 41055105 41055106 41055107 41055108 41055109 41055110 40105536 40105559 Display Section Display and OperationKeypad Programming and Calibrating the Solu Comp II Tutorial Using Hold SecuritySection Programming the Analyzer Alarm Section Default SettingsOutput Dampening Or 4 mA Mode Section Sensors Output SectionDefault Settings Definitions Assigning Outputs 1Choose 4-20 mA or 0-20 mA Choose Output ConfigureChoose Output1 or Output2 Choose Output Range. Choose Output1 or Output2Alarm Relays Must be Configured Before Assigning Setpoints Choose Alarm Configure Choose Measurement or TempChoose Alarm Setpoints Definitions pH/ORP Definitions conductivity/resistivity Choose Measurement Choose Soln Temp Corr or Sensor IsoptntlDefinitions pH OutputsAlarms Measurement Choose Noise Rejection Choose ResetAnalyzer Procedure Choosing a Language Choose Default DisplayProcedure Changing Screen Contrast Section Calibration IntroductionCalibrating Temperature Procedure PH at 25C Standards Nominal pH Calibration Slope and OffsetChoose Measurement Choose BufferCal EnterManual Calibration pH S1ManualCal? Standardization pH Choose Measurement Choose StandardizeEntering a Known Slope Value pH Preparation of ORP standard solutions ORP CalibrationCalibrating AN IN-SERVICE Sensor Conductivity Procedure-Calibration against a standard meter and cellChoose Measurement Choose InProcess Procedure-Calibration against a standard solution ProgramDisplay Calibrate? Placing a NEW Sensor in Service Conductivity Choose Measurement Choose CellConstCalibrating the Analyzer Conductivity Cell constant Conductivity µS/cm Resistor k ΩTo Reset the Factory Resistance CALIBRATION, see Section Overview Section MaintenanceReplacement Parts Model Description See FigureExploded View of Solu Comp II Panel Mount Version Exploded View of Solu Comp II Pipe/Surface Mount Version Simulating Inputs Conductivity Simulating Temperature Section TroubleshootingOverview Troubleshooting Using Fault Codes Absolute Value of Measured Voltage Exceeds 2500 mVRTD Sense Line for Sensor 1 or Sensor 2 is Open PH Sensitive Glass Membrane is BrokenConductivity Sensor Open or Shorted Eeprom FailureProblem See Section Calibration Error During Two-Point CalibrationInvalid Input While Manually Entering Slope Calibration Error during StandardizationSensor Does Not Respond to Known pH Changes Model Solu Comp Section Troubleshooting Calibration Error Using Standard Resistors Conductivity, µS/cm Cell constant, /cm Alarm Relays Do Not WorkMeasurement seems wrong Current Output Too LowGeneral Simulating pH input when the preamplifier is in the sensorPreamplifier is in the Sensor Set the temperature correction to raw Section Simulating Inputs ConductivitySimulating temperature Three-Wire RTD ConfigurationMeasuring Reference Voltage Checking for a Poisoned Reference ElectrodeSection Return of Material Warranty RepairNON-WARRANTY Repair GeneralWarranty Specifications subject to change without notice

PN 51-1055pHC/rev.K specifications

The Emerson PN 51-1055pHC/rev.K is a highly advanced and reliable component designed for a wide range of applications in industrial automation and control systems. This device is known for its robust performance, exceptional durability, and integration of cutting-edge technologies, making it an ideal choice for demanding operational environments.

One of the key features of the PN 51-1055pHC/rev.K is its versatility. It is compatible with various systems and can be easily integrated into existing configurations without the need for extensive modifications. This adaptability allows users to streamline their operations and enhance overall efficiency, ensuring that the PN 51-1055pHC/rev.K is suitable for diverse industrial scenarios.

Technologically, the device incorporates advanced sensor capabilities that enable accurate monitoring and measurement of critical parameters. These sensors are designed to provide real-time data, allowing operators to make informed decisions quickly. The real-time diagnostics feature is essential for maintenance, as it helps predict potential issues before they lead to system failures, thereby reducing downtime and operational costs.

The PN 51-1055pHC/rev.K also boasts a rugged construction, built to withstand harsh industrial conditions such as extreme temperatures, vibrations, and moisture. This durability ensures a long lifespan, making it a cost-effective investment for businesses looking to optimize their operations without frequent replacements or repairs.

Another critical characteristic of the Emerson PN 51-1055pHC/rev.K is its user-friendly interface. The device is designed with ease of use in mind, featuring intuitive controls and straightforward setup processes. This accessibility enhances user experience, allowing both seasoned technicians and new operators to utilize its capabilities effectively.

Additionally, the device supports various communication protocols, enabling seamless integration with other equipment and systems. This feature is crucial for creating interconnected environments where data can be shared and analyzed across multiple platforms, fostering smarter and more efficient industrial processes.

In summary, the Emerson PN 51-1055pHC/rev.K stands out for its versatility, advanced sensor technologies, rugged design, user-friendly interface, and communication capabilities. It is a powerful tool for businesses seeking improved automation and control, helping to drive efficiency and enhance operational reliability in any industrial setting.