Emerson PN 51-1055pHC/rev.K Problem See Section, Calibration Error During Two-Point Calibration

Page 76

MODEL SOLU COMP II

SECTION 8.0

 

TROUBLESHOOTING

8.3 TROUBLESHOOTING WHEN NO ERROR MESSAGE IS SHOWING — pH.

 

Problem

See Section

 

 

 

 

 

 

New temperature during calibration more than 2-3°C different from the live reading

8.3.1

 

 

 

 

 

 

Calibration Error warning during two-point calibration

8.3.2

 

 

 

 

 

 

Calibration Error warning during standardization

8.3.3

 

 

 

 

 

 

Invalid Input while manually entering slope

8.3.4

 

 

 

 

 

 

Sensor does not respond to known pH changes

8.3.5

 

 

Calibration was successful, but process pH is slightly different from expected value

8.3.6

 

 

 

 

 

 

Calibration was successful, but process pH is grossly wrong and/or noisy

8.3.7

 

 

 

 

 

8.3.1 Difference Between Solu Comp II and Standard Thermometer is Greater Than 3°C.

A.Is the standard thermometer, RTD, or thermistor accurate? General purpose liquid-in-glass thermometers, par- ticularly ones that have been mistreated, can have surprisingly large errors.

B.Is the temperature element in the pH sensor completely submerged in the test liquid?

C.Is the standard temperature sensor submerged to the correct level?

D.Review Section 6.2.

8.3.2 Calibration Error During Two-Point Calibration

Once the two-point (manual or automatic) calibration is complete, the Solu Comp II automatically calculates the sensor slope (at 25°). If the slope is greater than 60 mV/pH or less than 45 mV/pH, the analyzer displays the Calibration Error screen and does not update the calibration. Check the following:

A.Are the buffers accurate? Inspect the buffers for obvious signs of deterioration, such as turbidity or mold growth. Neutral and slightly acidic buffers are highly susceptible to molds. Alkaline buffers (pH 9 and greater) that have been exposed to air for long periods may also be inaccurate. Alkaline buffers absorb carbon dioxide from the atmosphere, which lowers the pH. If a high pH buffer was used in the failed calibration, repeat the calibration using a fresh buffer. If fresh buffer is not available, use a lower pH buffer. For example, use pH 4 and 7 buffer instead of pH 7 and 10 buffer.

B.Was adequate time allowed for temperature equilibration? If the sensor was in a process substantially hotter or colder than the buffer, place it in a container of water at ambient temperature for at least 20 minutes before starting the calibration. Using auto calibration avoids calibration errors caused by temperature drift. The ana- lyzer will not update readings until the drift is less than 0.02 pH over 10 seconds.

C.Were correct pH values entered during manual calibration? Using auto calibration eliminates errors caused by improperly entering data.

D.Is the sensor properly wired to the analyzer? Check the sensor wiring including any connections in a junction box. See Section 3.2.

E.Is the sensor dirty or coated? See the sensor instruction manual for cleaning instructions.

SRef Imp

123k^

Glass Imp

123M^

 

 

F. Is the sensor faulty? With the main display showing, use or to scroll through the information screens until the electrode impedance screen (at left) is displayed. Refer to the table on the following page for an interpretation of the impedance readings.

Another way of checking for a faulty sensor is to replace it with a new one. If the new sensor can be calibrat- ed, the old sensor has failed.

G. Is the analyzer faulty? The best way to check for a faulty analyzer is to simulate pH inputs. See Section 8.6.

67

Image 76
Contents Model Solu Comp Essential Instructions Quick Start Guide ORP Quick Reference Guide Model Solu Comp Table of Contents CONT’D List of TablesList of Figures IiiAbout This Document Section Description and Specifications Dimensions Terminal Connections Rating 26-14 AWG wire sizeHazardous Location Ordinary Location -68 onlyPerformance Specifications Analyzer Conductivity InputRecommended Sensors for Conductivity Range, µS/cm Cell constant, /cmToroidal Conductivity Codes -21 and/or Measurement Range see table belowPerformance Specifications Analyzer ORP Input Temperature Correction none requiredWhen used with conductivity -20-32 or Recommended Sensors for ORPFrequency Range 0.5 4000 Hz Recommended SensorsRepeatability ±1% Loop Specifications with a Model 499A OZ Sensor Recommended SensorInput filter 1- 255 samples Measurement Range 0-10 ppm mg/LOrdering Information Model 1055 Solu Comp II AnalyzerCombination Guide Part Number Description Weight Section Installation InstallationUnpacking and Inspection General InformationPanel Mount Installation Panel MountingPipe Mount Installation Pipe MountingSurface Mount Installation Surface MountingPOWER, ALARM, OUTPUT, and Sensor Connections Section WiringPreparing Conduit Openings Model Solu Comp Section Wiring Model Solu Comp Section Wiring 41055105 41055106 41055107 41055108 41055109 41055110 40105536 40105559 Display Section Display and OperationKeypad Programming and Calibrating the Solu Comp II Tutorial Using Hold SecuritySection Programming the Analyzer Default Settings Output Dampening Or 4 mA Mode SectionAlarm Section Sensors Output SectionDefault Settings Definitions Assigning Outputs 1Choose Output Configure Choose Output1 or Output2Choose 4-20 mA or 0-20 mA Choose Output Range. Choose Output1 or Output2Alarm Relays Must be Configured Before Assigning Setpoints Choose Alarm Configure Choose Measurement or TempChoose Alarm Setpoints Definitions pH/ORP Definitions conductivity/resistivity Choose Measurement Choose Soln Temp Corr or Sensor IsoptntlDefinitions pH OutputsAlarms Measurement Choose Noise Rejection Choose ResetAnalyzer Procedure Choosing a Language Choose Default DisplayProcedure Changing Screen Contrast Section Calibration IntroductionCalibrating Temperature Procedure PH at 25C Standards Nominal pH Calibration Slope and OffsetChoose Measurement Choose BufferCal EnterManual Calibration pH S1ManualCal? Standardization pH Choose Measurement Choose StandardizeEntering a Known Slope Value pH Preparation of ORP standard solutions ORP CalibrationCalibrating AN IN-SERVICE Sensor Conductivity Procedure-Calibration against a standard meter and cellChoose Measurement Choose InProcess Procedure-Calibration against a standard solution ProgramDisplay Calibrate? Placing a NEW Sensor in Service Conductivity Choose Measurement Choose CellConstCalibrating the Analyzer Conductivity Cell constant Conductivity µS/cm Resistor k ΩTo Reset the Factory Resistance CALIBRATION, see Section Section Maintenance Replacement PartsOverview Model Description See FigureExploded View of Solu Comp II Panel Mount Version Exploded View of Solu Comp II Pipe/Surface Mount Version Section Troubleshooting Overview Troubleshooting Using Fault CodesSimulating Inputs Conductivity Simulating Temperature Absolute Value of Measured Voltage Exceeds 2500 mVPH Sensitive Glass Membrane is Broken Conductivity Sensor Open or ShortedRTD Sense Line for Sensor 1 or Sensor 2 is Open Eeprom FailureProblem See Section Calibration Error During Two-Point CalibrationInvalid Input While Manually Entering Slope Calibration Error during StandardizationSensor Does Not Respond to Known pH Changes Model Solu Comp Section Troubleshooting Calibration Error Using Standard Resistors Alarm Relays Do Not Work Measurement seems wrongConductivity, µS/cm Cell constant, /cm Current Output Too LowGeneral Simulating pH input when the preamplifier is in the sensorPreamplifier is in the Sensor Set the temperature correction to raw Section Simulating Inputs ConductivitySimulating temperature Three-Wire RTD ConfigurationMeasuring Reference Voltage Checking for a Poisoned Reference ElectrodeWarranty Repair NON-WARRANTY RepairSection Return of Material GeneralWarranty Specifications subject to change without notice

PN 51-1055pHC/rev.K specifications

The Emerson PN 51-1055pHC/rev.K is a highly advanced and reliable component designed for a wide range of applications in industrial automation and control systems. This device is known for its robust performance, exceptional durability, and integration of cutting-edge technologies, making it an ideal choice for demanding operational environments.

One of the key features of the PN 51-1055pHC/rev.K is its versatility. It is compatible with various systems and can be easily integrated into existing configurations without the need for extensive modifications. This adaptability allows users to streamline their operations and enhance overall efficiency, ensuring that the PN 51-1055pHC/rev.K is suitable for diverse industrial scenarios.

Technologically, the device incorporates advanced sensor capabilities that enable accurate monitoring and measurement of critical parameters. These sensors are designed to provide real-time data, allowing operators to make informed decisions quickly. The real-time diagnostics feature is essential for maintenance, as it helps predict potential issues before they lead to system failures, thereby reducing downtime and operational costs.

The PN 51-1055pHC/rev.K also boasts a rugged construction, built to withstand harsh industrial conditions such as extreme temperatures, vibrations, and moisture. This durability ensures a long lifespan, making it a cost-effective investment for businesses looking to optimize their operations without frequent replacements or repairs.

Another critical characteristic of the Emerson PN 51-1055pHC/rev.K is its user-friendly interface. The device is designed with ease of use in mind, featuring intuitive controls and straightforward setup processes. This accessibility enhances user experience, allowing both seasoned technicians and new operators to utilize its capabilities effectively.

Additionally, the device supports various communication protocols, enabling seamless integration with other equipment and systems. This feature is crucial for creating interconnected environments where data can be shared and analyzed across multiple platforms, fostering smarter and more efficient industrial processes.

In summary, the Emerson PN 51-1055pHC/rev.K stands out for its versatility, advanced sensor technologies, rugged design, user-friendly interface, and communication capabilities. It is a powerful tool for businesses seeking improved automation and control, helping to drive efficiency and enhance operational reliability in any industrial setting.