Emerson PN 51-1055pHC/rev.K Dimensions, Hazardous Location, Ordinary Location -68 only

Page 11

MODEL SOLU COMP II

1.2 SPECIFICATIONS - General

Case: ABS (panel-mount), polycarbonate (pipe- and surface-mount). All versions are NEMA 4X/CSA 4 (IP65).

Dimensions

Panel (code -10):6.10 x 6.10 x 3.72 in. (155 x 155 x 94.5 mm)

Surface/Pipe (code -11): 6.23 x 6.23 x 3.23 in. (158 x 158 x 82 mm); see page 5 for dimensions of pipe mounting bracket.

Conduit openings: Accepts PG13.5 or 1/2 in. con- duit fittings

Display: Two line, 16-character, back-lit display. Character height: 4.8 mm. Display can be cus- tomized to meet individual requirements. Depending on number of sensors, as many as 14 display screens are available.

Ambient temperature and humidity: 0 to 50°C, (32 to 122°F) RH 5 to 95% (non-condensing)

Note: The analyzer is operable from -20 to 60°C (-4 to 140°F) with some degradation in display performance.

Power:

Code -01: 115/230 Vac ±15%, 50/60 Hz ±6%, 8.0W

Code -02*: 24 Vdc ±15%, 6.0W Installation Category II

*For +24Vdc Power Supply use only devices meeting NEC Class II or UL recognized (UL 1950).

Equipment protected throughout by double insula- tion.

Hazardous Location:

Class I, Division 2,

Groups A, B, C, & D

POLLUTION DEGREE 4: Extended Environment Outdoor use where conductive contamination such as rain, snow, or dust may be present. (Hazardous Location only)

Ordinary Location: (-68 only)

12RN

POLLUTION DEGREE 2: Normally only non-con- ductive pollution occurs. Occasionally, however, a temporary conductivity caused by condensation must be expected.

SECTION 1.0

DESCRIPTION AND SPECIFICATIONS

RFI/EMI: EN-61326

LVD: EN-61010-1

Input: Choice of single or dual sensor input with measurement choices of pH/ORP, conductivity/ resistivity, toroidal conductivity, flow, chlorine, dis- solved oxygen, and dissolved ozone. Field-com- missioned units allow user to change measure- ments on either or both inputs. See combination guide for valid combinations. For contacting con- ductivity measurements, temperature element must be a Pt 1000 RTD. For other measure- ments, use either a Pt100 RTD, Pt1000 RTD, or 22k NTC (D.O. only).

Outputs: Two 4-20 mA or 0-20 mA isolated outputs. Continuously adjustable. Linear or logarithmic. Maximum load 500 ohms. Output dampening with time constant of 5 sec is user-selectable.

Alarms: Three alarm relays for process measure- ment(s) or temperature. Alarm 3 can be configured as a fault alarm, instead of a process alarm. Each relay can be configured

independently. Alarm logic (high or low activation or USP*) and deadband are user-programmable. The USP* alarm can be programmed to activate when the conductivity is within a user-selectable percentage of the limit.

*conductivity/resistivity measurement only

Relays: Form C, single pole double throw, epoxy sealed

 

 

Resistive

Inductive

 

 

28 Vdc

5.0 A

3.0 A

 

 

115 Vac

5.0 A

3.0 A

 

 

230 Vac

5.0 A

1.5 A

 

Terminal Connections Rating: 26-14 AWG wire size

Weight/Shipping weight (rounded up to nearest lb or nearest 0.5 kg): 3 lb (1.5 kg)/4 lb (2.0 kg)

2

Image 11
Contents Model Solu Comp Essential Instructions Quick Start Guide ORP Quick Reference Guide Model Solu Comp List of Tables Table of Contents CONT’DIii List of FiguresAbout This Document Section Description and Specifications Ordinary Location -68 only DimensionsTerminal Connections Rating 26-14 AWG wire size Hazardous LocationRange, µS/cm Cell constant, /cm Performance SpecificationsAnalyzer Conductivity Input Recommended Sensors for ConductivityMeasurement Range see table below Toroidal Conductivity Codes -21 and/orRecommended Sensors for ORP Performance Specifications Analyzer ORP InputTemperature Correction none required When used with conductivity -20-32 orRepeatability ±1% Recommended SensorsFrequency Range 0.5 4000 Hz Measurement Range 0-10 ppm mg/L Loop Specifications with a Model 499A OZ SensorRecommended Sensor Input filter 1- 255 samplesModel 1055 Solu Comp II Analyzer Ordering InformationCombination Guide Part Number Description Weight General Information Section InstallationInstallation Unpacking and InspectionPanel Mounting Panel Mount InstallationPipe Mounting Pipe Mount InstallationSurface Mounting Surface Mount InstallationPreparing Conduit Openings Section WiringPOWER, ALARM, OUTPUT, and Sensor Connections Model Solu Comp Section Wiring Model Solu Comp Section Wiring 41055105 41055106 41055107 41055108 41055109 41055110 40105536 40105559 Keypad Section Display and OperationDisplay Programming and Calibrating the Solu Comp II Tutorial Security Using HoldSection Programming the Analyzer Sensors Output Section Default SettingsOutput Dampening Or 4 mA Mode Section Alarm SectionDefault Settings Assigning Outputs 1 DefinitionsChoose Output Range. Choose Output1 or Output2 Choose Output ConfigureChoose Output1 or Output2 Choose 4-20 mA or 0-20 mAAlarm Relays Must be Configured Before Assigning Setpoints Choose Measurement or Temp Choose Alarm ConfigureChoose Alarm Setpoints Definitions pH/ORP Definitions conductivity/resistivity Choose Soln Temp Corr or Sensor Isoptntl Choose MeasurementDefinitions pH OutputsAlarms Measurement Choose Noise Rejection Choose ResetAnalyzer Procedure Changing Screen Contrast Choose Default DisplayProcedure Choosing a Language Introduction Section CalibrationCalibrating Temperature Procedure Calibration Slope and Offset PH at 25C Standards Nominal pHEnter Choose Measurement Choose BufferCalManual Calibration pH S1ManualCal? Choose Measurement Choose Standardize Standardization pHEntering a Known Slope Value pH ORP Calibration Preparation of ORP standard solutionsProcedure-Calibration against a standard meter and cell Calibrating AN IN-SERVICE Sensor ConductivityChoose Measurement Choose InProcess Procedure-Calibration against a standard solution ProgramDisplay Calibrate? Choose Measurement Choose CellConst Placing a NEW Sensor in Service ConductivityCell constant Conductivity µS/cm Resistor k Ω Calibrating the Analyzer ConductivityTo Reset the Factory Resistance CALIBRATION, see Section Model Description See Figure Section MaintenanceReplacement Parts OverviewExploded View of Solu Comp II Panel Mount Version Exploded View of Solu Comp II Pipe/Surface Mount Version Absolute Value of Measured Voltage Exceeds 2500 mV Section TroubleshootingOverview Troubleshooting Using Fault Codes Simulating Inputs Conductivity Simulating TemperatureEeprom Failure PH Sensitive Glass Membrane is BrokenConductivity Sensor Open or Shorted RTD Sense Line for Sensor 1 or Sensor 2 is OpenCalibration Error During Two-Point Calibration Problem See SectionSensor Does Not Respond to Known pH Changes Calibration Error during StandardizationInvalid Input While Manually Entering Slope Model Solu Comp Section Troubleshooting Calibration Error Using Standard Resistors Current Output Too Low Alarm Relays Do Not WorkMeasurement seems wrong Conductivity, µS/cm Cell constant, /cmPreamplifier is in the Sensor Simulating pH input when the preamplifier is in the sensorGeneral Simulating Inputs Conductivity Set the temperature correction to raw SectionThree-Wire RTD Configuration Simulating temperatureChecking for a Poisoned Reference Electrode Measuring Reference VoltageGeneral Warranty RepairNON-WARRANTY Repair Section Return of MaterialWarranty Specifications subject to change without notice

PN 51-1055pHC/rev.K specifications

The Emerson PN 51-1055pHC/rev.K is a highly advanced and reliable component designed for a wide range of applications in industrial automation and control systems. This device is known for its robust performance, exceptional durability, and integration of cutting-edge technologies, making it an ideal choice for demanding operational environments.

One of the key features of the PN 51-1055pHC/rev.K is its versatility. It is compatible with various systems and can be easily integrated into existing configurations without the need for extensive modifications. This adaptability allows users to streamline their operations and enhance overall efficiency, ensuring that the PN 51-1055pHC/rev.K is suitable for diverse industrial scenarios.

Technologically, the device incorporates advanced sensor capabilities that enable accurate monitoring and measurement of critical parameters. These sensors are designed to provide real-time data, allowing operators to make informed decisions quickly. The real-time diagnostics feature is essential for maintenance, as it helps predict potential issues before they lead to system failures, thereby reducing downtime and operational costs.

The PN 51-1055pHC/rev.K also boasts a rugged construction, built to withstand harsh industrial conditions such as extreme temperatures, vibrations, and moisture. This durability ensures a long lifespan, making it a cost-effective investment for businesses looking to optimize their operations without frequent replacements or repairs.

Another critical characteristic of the Emerson PN 51-1055pHC/rev.K is its user-friendly interface. The device is designed with ease of use in mind, featuring intuitive controls and straightforward setup processes. This accessibility enhances user experience, allowing both seasoned technicians and new operators to utilize its capabilities effectively.

Additionally, the device supports various communication protocols, enabling seamless integration with other equipment and systems. This feature is crucial for creating interconnected environments where data can be shared and analyzed across multiple platforms, fostering smarter and more efficient industrial processes.

In summary, the Emerson PN 51-1055pHC/rev.K stands out for its versatility, advanced sensor technologies, rugged design, user-friendly interface, and communication capabilities. It is a powerful tool for businesses seeking improved automation and control, helping to drive efficiency and enhance operational reliability in any industrial setting.