Omega Engineering CCT-55-I manual Frequency Output Range Selection, Pulses PER Second

Page 7

FREQUENCY OUTPUT RANGE SELECTION

PULSES PER SECOND

Table 2 indicates all the standard available ranges for these models when the frequency output must be in pulses per second.

Set the jumpers/solder pads to the position indicated in Table 2. The jumpers are located on the "MS-55" output module.

If the frequency output range is modified follow the adjustment and calibration procedure for each model.

 

PULSES / SECOND

 

CLOSE JUMPERS/

PULSE WIDTH

 

 

AND SOLDER PADS

 

 

 

 

 

 

 

 

 

 

 

0 to 5000/10000

 

E, 2 & 5

Variable

 

 

 

 

 

 

 

0 to

2500/5000

 

E, 1 & 5

 

 

0 to

1250/2500

 

E, A, 1 & 5

 

 

0 to

625/1250

 

E, B, 1 & 5

 

 

0 to

312.5/625

 

E, B, A& 5

50 % of the full cycle

 

 

 

 

0 to 156.3/312.5

 

E, C, 1 & 5

 

 

0 to

78.1/156.3

 

E, C, A, 1 & 5

 

 

0 to

39.1/78.1

 

E, C, B, 1 & 5

 

 

0 to

19.5/39.1

 

E, C, B, A, 1 & 5

 

 

0 to

9.77/19.5

 

E, D, 1 & 5

 

 

0 to

4.88/9.77

 

E, D, A, 1 & 5

 

 

0 to

2.44/4.88

 

E, D, B, 1 & 5

 

 

 

 

TABLE 2

The frequency output for each range is indicated : 0 to 2500/5000 for example.

This means that the minimum frequency output is 2500 Hz. and the maximum frequency output is 5000 Hz. Any frequency within the low and high value, can be obtained using the trimmer P3 located on the "MS-55" output module.

6

Image 7
Contents User’sGuide Omega.comTM Table of Contents Internal Overview Open the HousingProcedure Procedure for DisassemblyChanging the Power Supply Power Supply module MATypes of Pulses OverviewPulses PER Second Frequency Output Range SelectionPulses / Hour Close Jumpers Pulse Width Solder Pads Pulses PER HourAmplifier Gain Offset AdjustOutput CCT-55-V DC Voltage Integrator Electrical FeaturesGeneral Specifications InputAdjustment and Calibration Procedure Signal Input Range SelectionRange Selection Signal Input CCT-55-I DC Current Integrator Electrical FeaturesInput CCT-55-V Wiring ConnectionsDimensions Mechanical SpecificationsTechnical Data Page USAWARRANTY/DISCLAIMER Temperature