Omega Engineering CCT-55-I manual Offset Adjust, Amplifier Gain

Page 9

SIGNAL INPUT, module ME

OVERVIEW

This module together with the "MP" input board, performs all the input signal conditioning (see Fig. 2). This module contains the trimmers and jumpers for the amplifier gain and the low level output (offset).

The signal input connections are made at Terminals 3, 4, 5 and 6.

Each model may use some of these terminals, see the Connections Section.

FIG.7

 

1

 

P1

2

P2

 

 

P1 : Output zero adjustment.

P2 : Gain amplifier adjustment.

B

A

C

Solder pads on solder side

OFFSET ADJUST

Solder pad 1 if closed Solder pad 2 if closed Jumper A if closed

:Adjust the low range level of the output (Offset positive coarse).

:Adjust the low range level of the output (Offset negative coarse).

:Adjust the low range level of the output (Offset negative fine).

AMPLIFIER GAIN

Jumper B if closed Jumper C if closed Jumpers B and C opened

:Gain at maximum level

:Gain at medium level.

:Gain at minimum level.

8

Image 9
Contents User’sGuide Omega.comTM Table of Contents Procedure Open the HousingProcedure for Disassembly Internal OverviewChanging the Power Supply Power Supply module MA Types of Pulses OverviewPulses PER Second Frequency Output Range SelectionPulses / Hour Close Jumpers Pulse Width Solder Pads Pulses PER HourAmplifier Gain Offset AdjustGeneral Specifications CCT-55-V DC Voltage Integrator Electrical FeaturesInput OutputSignal Input Range Selection Adjustment and Calibration ProcedureRange Selection Signal Input CCT-55-I DC Current Integrator Electrical FeaturesInput CCT-55-V Wiring ConnectionsMechanical Specifications DimensionsTechnical Data Page USAWARRANTY/DISCLAIMER Temperature