Sanyo MPR-1013R, MPR-513R instruction manual Alarms & Safety Functions, Alarms and safety functions

Page 19

ALARMS & SAFETY FUNCTIONS

This unit has the alarms and safety functions shown in Table 5, and also self diagnostic functions.

Table 5. Alarms and safety functions

Alarm & Safety

Situation

Indication

High temperature

If the chamber temperature deviates

ALARM lamp is flashed.

from the set temperature +2oC or up to

Temperature indicator is

alarm

+14oC.

flashed.

 

If the chamber temperature deviates

 

Low temperature

from the set temperature -2oC or up to

ALARM lamp is flashed.

-14oC.

Temperature indicator is

alarm

 

flashed.

If the chamber temperature is lower

 

 

than 0oC.

 

Over-heat

When the chamber temp. is higher

 

than 28oC. (Reset when the chamber

-----

protection

temp. is lower than about 18oC)

 

Over-cooling

When the chamber temp. is lower than

 

-1oC. (Reset when the chamber

-----

protection

temp. is higher than 8oC)

 

Power failure alarm

When the power to the unit is

-----

disconnected.

 

 

Door alarm

When the door is open.

Door check lamp is lit.

 

 

 

Auto-return

When there is no key pressing in

Chamber temperature is

each setting mode for 90 seconds.

displayed.

 

Key lock

When the key lock is “ON”.

-----

 

 

 

 

If the thermal sensor is disconnected

ALARM lamp is flashed.

 

E01 and 50oC (or -50oC)

 

or short-circuited.

are displayed alternately.

 

 

Thermal sensor

If the defrost sensor 1 is disconnected

ALARM lamp is flashed.

E02 and chamber temp.

abnormality

or short-circuited.

are displayed alternately.

 

 

 

If the defrost sensor 2 is disconnected

ALARM lamp is flashed.

 

or short-circuited. (MPR-1013,1013R

E03 and chamber temp.

 

only)

are displayed alternately.

Memory back-up

When power failure or disconnected

-----

the power cord.

 

 

Buzzer

Intermittent tone with 15 minutes delay.

Intermittent tone with 15 minutes delay.

Intermittent tone

-----

-----

-----

Intermittent tone with 2 minutes delay.

-----

-----

Intermittent tone

Intermittent tone

Intermittent tone

-----

Safety operation

Remote alarm with 15 minutes delay.

Remote alarm with 15 minutes delay.

Remote alarm

Inside fan, drain pan heater, defrost heater

OFF

Compressor OFF

Remote alarm.

-----

Finishing of each setting mode.

Change of setting is disable.

Remote alarm. Operation by the defrost sensor

Remote alarm. Unit keeps running.

Remote alarm. Unit keeps running.

Set value memorized

Note:

The remote alarm is silenced by pressing the buzzer stop key (BZ) as the remote alarm is operated in conjunction with the buzzer, except for the power failure alarm.

When more than two alarm conditions occur simultaneously, the lowest number error code has priority on the error display.

After a power failure, the unit will resume operation with the set value that was in place before power failure occurred.

The over-cooling protection may activate when the set temperature is 2oC. In this case, the chamber temperature will rise to 8oC, but this is not malfunction.

18

Image 19
Contents MPR-513 MPR-513R MPR-1013 MPR-1013R Contents Symbol shows Precautions for Safe OperationPrecautions for Safe Operation Prevent electric shock or injury May result Page Circuit breaker Environmental ConditionsRefrigerator Components Door switch Space for automatic temperature recorderAccess port Light switchControl panel and keypad Installation site InstallationFix the unit InstallationGround earth Optional Components Automatic temperature recorderAttachment of recorder MTR-G04 Forced defrost START-UP of UnitDefrost Cycles Cycle defrostChamber temperature Temperature SettingKey lock function Alarm temperature setting Alarms and safety functions Alarms & Safety FunctionsConnection of remote alarm terminal Remote alarm terminalCleaning of cabinet Routine MaintenanceCleaning of condenser filter MPR-1013, 1013R Cleaning of evaporating tray Replacement of glow starter Replacement of fluorescent lampDisposal of Unit TroubleshootingSpecifications Performance Safety check sheet 7FB6P101277006 5 April

MPR-513, MPR-1013R, MPR-1013, MPR-513R specifications

Sanyo, a renowned name in the world of electronics, has made significant strides in the air conditioning sector with its advanced models such as the MPR-513R, MPR-1013, MPR-1013R, and MPR-513. These models integrate cutting-edge technologies and user-oriented features designed to enhance indoor comfort while promoting energy efficiency.

The Sanyo MPR-513R is celebrated for its powerful cooling and heating capabilities. It is designed for medium to large spaces, making it ideal for residential and commercial applications alike. This model leverages inverter technology, which allows for variable speed operation, ensuring that the air conditioning system adjusts seamlessly according to your needs. This not only enhances comfort but also significantly reduces energy consumption, making it an economical choice.

The MPR-1013 and MPR-1013R models follow closely in performance. These units are particularly known for their compact design and lightweight construction, making them easier to install in diverse environments. Both models boast impressive energy efficiency ratings and utilize eco-friendly refrigerants, supporting Sanyo's commitment to sustainability. Their quiet operation is another highlight, ensuring that users enjoy a comfortable environment without the distraction of loud machinery.

On the other hand, the MPR-513 model is distinguished by its advanced filtration system. It is equipped with a multi-stage filtration process that captures dust, allergens, and airborne particles, providing clean and healthy air. Ideal for allergy sufferers, this model emphasizes indoor air quality alongside cooling and heating performance.

All four models come with user-friendly controls, allowing for easy temperature adjustments, timer settings, and remote operation. This feature enhances convenience, enabling users to customize their indoor climate effortlessly.

Overall, the Sanyo MPR series epitomizes innovation with its blend of advanced technologies and thoughtful design. Built to cater to the needs of modern users, these air conditioning units deliver effective climate control while prioritizing energy-saving features and environmental responsibility. Whether for home or office use, the MPR-513R, MPR-1013, MPR-1013R, and MPR-513 models stand out as reliable choices in today's competitive air conditioning market.